Comparison of an automated chemiluminescent assay to a manual ELISA assay for determination of von Willebrand Factor collagen binding activity on VWD plasma patients previously diagnosed through molecular analysis of VWF.
Аннотация
INTRODUCTION
The correct diagnosis and classification of VWD (von Willebrand disease) is crucial and must be optimized by including the collagen binding assay (VWF:CB). VWF:CB remains an under-recognized tool, not fully automated. The objective of this study was to evaluate and to compare the previously evaluated automated chemiluminescent assay (HemosIL AcuStar VWF:CB) to the ELISA ASSERACHROM® assay used routinely in our laboratory in patients with molecular diagnosis of VWD.
METHODS
A plasma sample from 49 patients previously diagnosed with VWD (type 1; type 2A, type 2M, type 2B) through phenotype and VWF (von Willebrand factor) analysis and 15 healthy controls was analysed. The VWF ristocetin cofactor activity (VWF:Rco) and VWF antigen (VWF:Ag) were performed simultaneously on the VWD plasma samples, and VWF:CB/VWF:Ag ratios were calculated.
RESULTS
The AcuStar VWF:CB assay was quickly performed with Pearson's correlation coefficient (r²) of .9571 between assays and a bias of +5.1U/dL (AcuStar > ELISA). Discrepancies of VWF:CB/VWF:Ag ratio were observed in type 2M-2A-like VWD (ratio <0.6 with AcuStar assay in 4/5 samples).
CONCLUSION
The AcuStar VWF:CB assay has demonstrated good performance to detect VWF mutational changes with capacity to discriminate quickly principal types of VWD.