Comparative Study of Arctic Front Advance Pro and POLARx Cryoballoons for Linear Ablation of the Left Atrial Roof.
Аннотация
BACKGROUND
The effectiveness of cryoballoon ablation (CBA) of the left atrial (LA) roof in addition to pulmonary vein isolation (PVI) using a novel cryoballoon catheter, POLARx, remains unclear.
METHODS
This study compared the efficacy of LA roof line ablation and PVI using POLARx (Boston Scientific) or AFA-Pro (Medtronic) in 100 patients with persistent atrial fibrillation. The right superior pulmonary vein (PV) anchoring and raise-up techniques were consistently used for LA roof line ablation, and rapid right ventricular pacing was applied if the cryoballoon temperature did not reach -40°C.
RESULTS
Complete conduction block at the LA roof could be obtained in all patients with POLARx and in 98.0% of patients with AFA-Pro. Rapid right ventricular pacing was needed in 64.0% of patients with AFA-Pro and in no patients with POLARx. During LA roof line ablation, the nadir cryoballoon temperature was significantly lower with POLARx than with AFA-Pro (right: -54.2°C ± 4.4°C vs. -46.0°C ± 5.4°C; central: -56.8°C ± 4.4°C vs. -45.7°C ± 4.8°C; left: -56.1°C ± 4.3°C vs. -46.1°C ± 5.7°C), and the cryoballoon temperature reached -40°C earlier with POLARx than with AFA-Pro (right: 30.8 ± 7.4 s vs. 74.1 ± 37.7 s; central: 28.2 ± 5.2 s vs. 62.9 ± 30.9 s; left: 29.8 ± 5.8 s vs. 69.6 ± 40.7 s).
CONCLUSION
The nadir cryoballoon temperature with POLARx was approximately 10°C lower than with AFA-Pro, consistently dropping below -40°C during LA roof line CBA. Thus, a complete conduction block of the LA roof line can be easily accomplished using right superior PV anchoring and the raise-up techniques without the need for rapid right ventricular pacing with POLARx.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.