Initial real-world data on catheter ablation in patients with persistent atrial fibrillation using the novel lattice-tip focal pulsed-field ablation catheter.
Аннотация
AIMS: Technological advancements have contributed to the enhanced precision and lesion flexibility in pulsed-field ablation (PFA) by integrating a three-dimensional mapping system combined with a point-by-point ablation strategy. Data regarding the feasibility of this technology remain limited to some clinical trials. This study aims to elucidate initial real-world data on catheter ablation utilizing a lattice-tip focal PFA/radiofrequency ablation (RFA) catheter in patients with persistent atrial fibrillation (AF).
METHODS AND
RESULTS
Consecutive patients who underwent catheter ablation for persistent AF via the lattice-tip PFA/RFA catheter were enrolled. We evaluated acute procedural data including periprocedural data as well as the clinical follow-up within a 90-day blanking period. In total, 28 patients with persistent AF underwent AF ablation either under general anaesthesia (n = 6) or deep sedation (n = 22). In all patients, pulmonary vein isolation was successfully achieved. Additional linear ablations were conducted in 21 patients (78%) with a combination of successful anterior line (n = 13, 46%) and roof line (n = 19, 68%). The median procedural and fluoroscopic times were 97 (interquartile range, IQR: 80-114) min and 8.5 (IQR: 7.2-9.5) min, respectively. A total of 27 patients (96%) were interviewed during the follow-up within the blanking period, and early recurrent AF was documented in four patients (15%) including one case of recurrent AF during the hospital stay. Neither major nor minor procedural complication occurred.
CONCLUSION
In terms of real-world data, our data confirmed AF ablation feasibility utilizing the lattice-tip focal PFA/RFA catheter in patients with persistent AF.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.