Intensive care medicine · 2000
Comparison of invasive and noninvasive measurements of indocyanine green plasma disappearance rate in critically ill patients with mechanical ventilation and stable hemodynamics.
Сравнительное исследованиеСтатьяВалидационное исследованиеНазвание и аннотация: Перевод пока не подготовлен
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Sakka S G, Reinhart K, Meier-Hellmann A
Аннотация
OBJECTIVES: We studied the correlation between invasive (aortic fiberoptic) and noninvasive (transcutaneous sensor) measurements of indocyanine green (ICG) plasma disappearance rate (PDR) in critically ill patients.
DESIGN AND SETTING: Prospective clinical study in a surgical intensive care unit of a university hospital.
PATIENTS: 16 critically ill patients with adult respiratory distress syndrome (n = 8), sepsis/septic shock (n = 6), subarachnoid hemorrhage (n = 1), or severe head injury (n = 1).
MEASUREMENTS AND RESULTS: We analyzed 16 pairs of simultaneous ICG PDR measurements. All patients were deeply sedated and mechanically ventilated. Each patient received a 4-F aortic catheter with an integrated fiberoptic and thermistor connected to a computer system for automatic calculation of invasive ICG PDR (PDRINV). An ICG sensor was also attached to the nose wing and connected to a DDG2001 analyzer for noninvasive measurement (PDRNINV). Linear regression analysis revealed PDRNINV = 0.98 PDRINV +0.11%/min (r = 0.94, p < 0.0001) with a mean bias of 0.2 +/- 2.0%/min.
CONCLUSION: Noninvasive measurements of ICG PDR are very highly correlated with values derived from an invasive fiberoptic-based reference technique.
DESIGN AND SETTING: Prospective clinical study in a surgical intensive care unit of a university hospital.
PATIENTS: 16 critically ill patients with adult respiratory distress syndrome (n = 8), sepsis/septic shock (n = 6), subarachnoid hemorrhage (n = 1), or severe head injury (n = 1).
MEASUREMENTS AND RESULTS: We analyzed 16 pairs of simultaneous ICG PDR measurements. All patients were deeply sedated and mechanically ventilated. Each patient received a 4-F aortic catheter with an integrated fiberoptic and thermistor connected to a computer system for automatic calculation of invasive ICG PDR (PDRINV). An ICG sensor was also attached to the nose wing and connected to a DDG2001 analyzer for noninvasive measurement (PDRNINV). Linear regression analysis revealed PDRNINV = 0.98 PDRINV +0.11%/min (r = 0.94, p < 0.0001) with a mean bias of 0.2 +/- 2.0%/min.
CONCLUSION: Noninvasive measurements of ICG PDR are very highly correlated with values derived from an invasive fiberoptic-based reference technique.
Предметные рубрики MeSH
AdolescentAdultAgedAged, 80 and overBiasBlood Gas Monitoring, TranscutaneousColoring AgentsCraniocerebral TraumaCritical IllnessDye Dilution TechniqueFemaleHemodynamicsHumansIndocyanine GreenLinear ModelsLiver Function TestsMaleMetabolic Clearance RateMiddle AgedMonitoring, PhysiologicProspective StudiesRespiration, ArtificialRespiratory Distress SyndromeSepsisSubarachnoid Hemorrhage