A novel double-balloon catheter for percutaneous balloon pulmonary valvuloplasty under echocardiographic guidance only.

Deng Run-Di, Zhang Feng-Wen, Zhao Guang-Zhi, Wen Bin, Wang Shou-Zheng, Ou-Yang Wen-Bin, Liu Yao, Xie Yong-Quan, Pan Xiang-Bin

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Percutaneous balloon pulmonary valvuloplasty (PBPV) is the procedure of choice for uncomplicated severe or symptomatic pulmonary stenosis. Echocardiography (echo)-guided PBPV can completely avoid the use of radiation and contrast agents compared to fluoroscopy-guided PBPV. Although we have confirmed that echo-guided PBPV is feasible in humans, the poor visibility of the traditional catheter under echo greatly limits the promotion of this new technology.

METHODS

We produced a novel double-balloon catheter to make the catheter easy to be detected by echo through adding a guiding balloon at the distal end of the catheter. Echo-guided PBPV was performed on thirty healthy swine using either a novel catheter or a traditional catheter to evaluate the feasibility and safety of the novel double-balloon catheter. The feasibility was evaluated by the success rate of balloon inflation at the pulmonary valve annulus and the operating time. The safety was evaluated by the frequency of balloon slippage and the incidence of complications.

RESULTS

There were no significant between-group differences in terms of weight and the ratio of balloon diameter to pulmonary annulus diameter. The success rate was 93.3% and 60% in the novel and traditional groups, respectively. The novel group had significantly (p<0.05) lower mean procedure time (6.33±6.86min vs 24.8±9.79min) and lower frequency of balloon slippage (0.07±0.26 vs 0.53±0.52), arrhythmia (0.07±0.26 vs 0.47±0.52), and tricuspid regurgitation (6.7% vs 40%) than the traditional group. No myocardial hematoma or pericardial tamponade occurred in the novel catheter group.

CONCLUSION

Although further studies and improvements are required, the study results indicate that the novel double-balloon catheter for echo-guided PBPV is feasible and safe.

Для цитирования: Deng Run-Di, Zhang Feng-Wen, Zhao Guang-Zhi, Wen Bin, Wang Shou-Zheng, Ou-Yang Wen-Bin, Liu Yao, Xie Yong-Quan, Pan Xiang-Bin. A novel double-balloon catheter for percutaneous balloon pulmonary valvuloplasty under echocardiographic guidance only.. Journal of cardiology. 2020. DOI: 10.1016/j.jjcc.2020.03.014.