DOI 10.1097/j.jcrs.0000000000001712

Медицинская библиотека · Journal of cataract and refractive surgery · 2025

Comparison of an image-guided system with four other devices for planning precise toric intraocular lens power calculation and axis alignment.

Melianets Volha, Kaiser Klemens Paul, Hemkeppler Eva, Kohnen Thomas, Böhm Myriam

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: To compare astigmatism and other keratometric parameters obtained with IOLMaster 500/700, Pentacam HR/AXL, and the VERION Image-Guided System.

SETTING

Department of Ophthalmology, Goethe-University, Frankfurt am Main, Germany.

DESIGN

Retrospective, comparative clinical study.

METHODS

1107 eyes of 654 patients were examined. The outcome parameters analyzed included the horizontal corneal diameter (white-to-white [WTW]), flattest (K1) and steepest keratometry (K2), axis values (K1 and K2 axis), astigmatism, and axis of astigmatism (axis). These parameters were compared between devices using the Bland-Altman method.

RESULTS

There was no significant difference between the VERION and the IOLMaster 500/700 in corneal astigmatism measurements ( P > .05). The astigmatism measurements with the Pentacam HR/AXL showed a statistically significant difference compared with the VERION Image-Guided System. The astigmatism measured with the VERION Image-Guided System was -0.04 ± 0.38 mm smaller ( P = .037) when compared with the Pentacam HR and -0.07 ± 0.37 mm smaller ( P < .001) than with the Pentacam AXL. The measurements of the astigmatic axis, as well as the K1 axis and the K2 axis showed a high agreement for all devices ( P > .05). However, significant differences were found between the devices for the WTW, K1, and K2 measurements ( P < .05).

CONCLUSIONS

The astigmatism measured by the VERION Image-Guided System is comparable with the measurements obtained by IOLMaster 500/700 and Pentacam HR/AXL, allowing the devices to be used interchangeably when planning astigmatism correction during cataract and refractive surgery. By limiting the use to measurements in clinic with "green" quality rating, a high agreement for all variables between the VERION Image-Guided System and IOLMaster 500/700 can be achieved.

Для цитирования: Melianets Volha, Kaiser Klemens Paul, Hemkeppler Eva, Kohnen Thomas, Böhm Myriam. Comparison of an image-guided system with four other devices for planning precise toric intraocular lens power calculation and axis alignment.. Journal of cataract and refractive surgery. 2025. DOI: 10.1097/j.jcrs.0000000000001712.