DOI 10.1080/08164622.2025.2486688

Медицинская библиотека · Clinical & experimental optometry · 2025

Repeatability and agreement of three different ocular biometers including single and dual Scheimpflug cameras.

Lwowski Christoph, Kaiser Klemens Paul, Ehrdorf Kevin, Shajari Mehdi, Kohnen Thomas

СтатьяСравнительное исследованиеПеревод пока не подготовлен

Аннотация

CLINICAL RELEVANCE: With a wide range of commercially available biometry devices, direct comparisons for agreement, interchangeability and repeatability are needed to enable correct interpretation of measured values.

BACKGROUND

The purpose of this study was to evaluate the intra-device repeatability and inter-device agreement of three different ocular biometers.

METHODS

This is a prospective, comparative study conducted at the University Hospital, Goethe University, Frankfurt, Germany. Included were eyes with a cataract which showed no other pathology. One eye per patient was randomly selected and 3 consecutive measurements were taken with the IOLMaster 700, Galilei G6, and Pentacam AXL. Repeatability and agreement were evaluated for simulated keratometry (simK), astigmatism (simA) and its axis, total keratometry (TC), axial length (AL), anterior chamber depth (ACD), and pachymetry (CT).

RESULTS

simK, TC, AL, and CT show very high repeatability with an intra-class correlation (ICC) of > 0.9 and a COV < 0.5 each. Means of the measurements were significantly different between the devices for simK, AL, and TC (p < 0.05). ACD and simA also showed a high ICC, but a lower agreement between the devices. Fixed biases shown by Bland-Altman plots for simK seem to indicate that the values are not interchangeable.

CONCLUSION

Even though very high repeatability and agreement for most of the measured variables was observed with all three devices, fixed differences indicate that measured values are not completely interchangeable between the devices.

Для цитирования: Lwowski Christoph, Kaiser Klemens Paul, Ehrdorf Kevin, Shajari Mehdi, Kohnen Thomas. Repeatability and agreement of three different ocular biometers including single and dual Scheimpflug cameras.. Clinical & experimental optometry. 2025. DOI: 10.1080/08164622.2025.2486688.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.