Performance of a photoscreener in detecting accommodation spasm.
Аннотация
CLINICAL RELEVANCE: For a correct spectacle prescription, accommodation spasm should not be overlooked in the measurements made using an autorefractometer. Although this can be easily understood with cycloplegia, it is not practical. Therefore, investigating practical methods that can predict excess accommodation will be useful in clinical practice.
BACKGROUND
To evaluate the performance of a photoscreener in detecting pseudo-myopia due to accommodation spasm and whether it eliminates the need for cycloplegia.
METHODS
Study was conducted in Van Training and Research Hospital. The manifest refraction of the patients was measured using an by an autorefractometer followed by a photoscreener before cycloplegia. Cycloplegic refraction errors were detected using an autorefractometer. Patients with pseudo-myopia due to accommodation spasm were diagnosed by observing a significant hyperopic shift in cycloplegic autorefractometer values compared to the non-cycloplegic values and were included in study. Cycloplegic autorefractometer and non-cycloplegic photoscreener measurements were compared.
RESULTS
The mean non-cycloplegic autorefractometer values were as follows: --6.34 ± 1.29D, -0.42 ± 0.58D, and -6.55 ± 1.36D for spherical, cylindrical, and spherical equivalents, respectively. These values were 0.23 ± 0.63D, -0.21 ± 0.41D, and 0.11 ± 0.73D for cycloplegic autorefractometer measurements, respectively. Cycloplegic autorefractometer values showed a significant hyperopic shift of 6.57 ± 1.26D and 6.67 ± 1.34D for spherical and spherical equivalents, respectively, compared to non-cycloplegic values (p < 0.001 for both). There was no significant difference between non-cycloplegic photoscreener and cycloplegic autorefractometer measurements for spherical, cylindrical, and spherical equivalents (p = 0.126, p = 0.078, and p = 0.053, respectively). . There was agreement between the measurements in the Bland-Altman graph.
CONCLUSION
There was consistency between non-cycloplegic photoscreener and cycloplegic autorefractometer measurements in patients with pseudomyopia due to accommodation spasm. Therefore, photoscreener measurements can be a fast, practical, cost-effective examination method for screening these patients without the need for cycloplegia.