Bachman Daniel R, Park Sangeun, Thaveepunsan Sutee, Fitzsimmons James S, An Kai-Nan, OʼDriscoll Shawn W
Аннотация
OBJECTIVE: To evaluate the effect of intentional undersizing of prosthetic radial head implant diameters on joint contact pressures.
METHODS: Eight fresh-frozen cadaveric elbows were aligned in neutral extension and loaded with 100 N using a custom testing apparatus. Radiocapitellar contact pressures were recorded using a Tekscan thin-film pressure sensor. Prosthetic radial head replacement was performed with 2 prostheses: the Anatomic Radial Head and the Evolve Proline Radial Head prostheses. Each design was sized according to the manufacturer's recommendations and then again using 2-mm smaller radial heads.
RESULTS: Average and peak pressures were significantly higher with the Evolve than the Anatomic prostheses (P < 0.03 and 0.02, respectively). Peak pressures decreased from 4.2 ± 0.5 MPa to 2.9 ± 0.3 MPa for the Anatomic Radial Heads and from 5.6 ± 0.5 MPa to 3.9 ± 0.6 MPa when the Evolve Radial Heads were undersized by 2 mm. The mean pressures of the Anatomic Radial Heads (1.4 ± 0.1 MPa) did not change significantly with undersizing (1.3 ± 0.1 MPa, P = 0.12), whereas the mean pressures of the Evolve Radial Heads (1.6 ± 0.1 MPa) were significantly reduced with undersizing (1.4 ± 0.1 MPa, P < 0.02).
CONCLUSION: Both mean and peak pressures were initially high for the Evolve Radial Head sized based on the short axis diameter and were improved with further undersizing by 2 mm. Peak, but not mean, contact pressures were improved by undersizing the Anatomic prosthesis based on the long axis diameter.
CLINICAL RELEVANCE: These findings support the clinical recommendation of some surgeons to undersize the Evolve prosthesis by 2-mm smaller diameter than the current manufacturer's suggestion and give reason to consider doing the same for the Anatomic prosthesis.
METHODS: Eight fresh-frozen cadaveric elbows were aligned in neutral extension and loaded with 100 N using a custom testing apparatus. Radiocapitellar contact pressures were recorded using a Tekscan thin-film pressure sensor. Prosthetic radial head replacement was performed with 2 prostheses: the Anatomic Radial Head and the Evolve Proline Radial Head prostheses. Each design was sized according to the manufacturer's recommendations and then again using 2-mm smaller radial heads.
RESULTS: Average and peak pressures were significantly higher with the Evolve than the Anatomic prostheses (P < 0.03 and 0.02, respectively). Peak pressures decreased from 4.2 ± 0.5 MPa to 2.9 ± 0.3 MPa for the Anatomic Radial Heads and from 5.6 ± 0.5 MPa to 3.9 ± 0.6 MPa when the Evolve Radial Heads were undersized by 2 mm. The mean pressures of the Anatomic Radial Heads (1.4 ± 0.1 MPa) did not change significantly with undersizing (1.3 ± 0.1 MPa, P = 0.12), whereas the mean pressures of the Evolve Radial Heads (1.6 ± 0.1 MPa) were significantly reduced with undersizing (1.4 ± 0.1 MPa, P < 0.02).
CONCLUSION: Both mean and peak pressures were initially high for the Evolve Radial Head sized based on the short axis diameter and were improved with further undersizing by 2 mm. Peak, but not mean, contact pressures were improved by undersizing the Anatomic prosthesis based on the long axis diameter.
CLINICAL RELEVANCE: These findings support the clinical recommendation of some surgeons to undersize the Evolve prosthesis by 2-mm smaller diameter than the current manufacturer's suggestion and give reason to consider doing the same for the Anatomic prosthesis.