Double potential interval and transisthmus conduction time for prediction of cavotricuspid isthmus block after ablation of typical atrial flutter.

Tai Ching-Tai, Haque Azizul, Lin Yung-Kuo, Tsao Hsuan-Ming, Ding Yu-An, Chang Mau-Song, Chen Shih-Ann

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

Complete bi-directional isthmus block is the endpoint of typical atrial flutter ablation. The purpose of this study was to investigate the feasibility of the local double potential (DP) interval and the change in transisthmus conduction time for predicting complete isthmus block after ablation of the cavotricuspid isthmus.

METHODS

The study population consisted of 32 patients with typical atrial flutter after a procedure of radiofrequency (RF) ablation of the cavotricuspid isthmus (16 had incomplete block and 16 had complete block). The transisthmus conduction time was determined during pacing from the proximal coronary sinus and low lateral right atrium before and after RF ablation. The DP interval close to the ablation line was evaluated after final RF energy application.

RESULTS

In the counterclockwise direction, transisthmus conduction time had an increase of 37 +/- 25.4% and 127.3 +/- 35.5% (P < 0.001), and the DP interval was 63.3 +/- 8.7 ms and 120 +/- 17.4 ms (P < 0.001) after achievement of incomplete and complete block, respectively. The sensitivity, specificity, positive and negative predictive values of an increase in the transisthmus conduction time > or =50% were 100%, 81%, 84% and 100%, respectively; those of DP interval > or =100 ms were 100%. In the clockwise direction, transisthmus conduction time had an increase of 38.8 +/- 28.6% and 135.7 +/- 63.6% (P < 0.001), and the DP interval was 63.6 +/- 13.8 ms and 127.7 +/- 27.1 ms (P < 0.001) after achievement of incomplete and complete block, respectively. The sensitivity, specificity, positive and negative predictive values of an increase in the transisthmus conduction time > or =50% were 100%, 67%, 83% and 100%, respectively; those of the DP interval > or =100 ms were 100%.

CONCLUSIONS

The transisthmus conduction time > or =50% increase or DP interval > or =100 ms was feasible to predict complete bi-directional isthmus block.

Для цитирования: Tai Ching-Tai, Haque Azizul, Lin Yung-Kuo, Tsao Hsuan-Ming, Ding Yu-An, Chang Mau-Song, Chen Shih-Ann. Double potential interval and transisthmus conduction time for prediction of cavotricuspid isthmus block after ablation of typical atrial flutter.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2002. DOI: 10.1023/a:1020876317859.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.