DOI 10.1097/mao.0000000000003623
(Even Off-Label) Cochlear Implantation in Single-Sided Deafness and Asymmetric Hearing Loss Results in Measurable Objective and Subjective Benefit.
Аннотация
OBJECTIVE
To quantify objective and subjective outcomes in cochlear implant (CI) recipients with asymmetric hearing loss, including single-sided deafness (SSD) whose candidacy was determined on an ear-specific basis when word recognition was 50% or less.
STUDY
DESIGN
Retrospective case review.
SETTING
Tertiary referral center.
BACKGROUND
The effectiveness of CI in cases of SSD and asymmetric hearing loss (AHL) has been described in terms of tinnitus suppression, improved speech recognition in quiet and noise, enhanced localization ability, and improved quality of life. However, CI is not yet routinely offered as a top option or standard of care for these individuals. Recent Food and Drug Administration (FDA) labeling limits aided word recognition in the ear to be implanted to only 5% in cases of AHL/SSD, which is significantly poorer than 40 to 50%, which is often referenced in cases of bilateral hearing loss. Anecdotal experience suggests that patients with much better preoperative word recognition than 5% can benefit from CI.
METHODS
We conducted a retrospective chart review of all adult CI candidates, with one ear exhibiting consonant-nucleus-consonant (CNC) word recognition scores at least 50% and one ear not meeting CI candidacy (i.e., CNC word recognition >50%). Outcome variables of interest included word and sentence recognition and subjective handicap questionnaires (hearing, tinnitus, dizziness) and the Speech Spatial Qualities questionnaire.
RESULTS
Statistically and clinically significant improvement in speech understanding (word, sentence, sentence in noise) was noted for both the SSD and AHL groups in the implanted ear. There were statistically and clinically significant subjective improvements noted for both groups on the Hearing Handicap Inventory, the Tinnitus Handicap Inventory, and the Speech Spatial Qualities questionnaire by 1 month after activation. There were no significant differences between the AHL and SSD groups on either objective or subjective measures of the implanted ear. Individual word understanding improved for the majority of recipients across both groups and is not dependent on meeting the FDA criteria of less than 5%.
CONCLUSIONS
Cochlear implantation is a viable option with measurable objective and perceived benefits for recipients with preoperative aided CNC word scores exceeding current FDA labeling. There is no significant difference between the AHL and SSD groups, suggesting that candidacy and outcome expectations should be set based on the ear to be implanted alone, without regard for the ability of the better hearing ear.