MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/aos.15747

Медицинская библиотека · Acta ophthalmologica · 2023

Evaluation of a retinal oximetry image quality indicator in patients with cataract.

Klefter Oliver Niels, Erichsen Jesper Høiberg, Hansen Mathias Møller, Holm Lars Morten, Hardarson Sveinn Hakon, Stefánsson Einar, Kessel Line

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To evaluate a new automated retinal oximetry image quality indicator with cataract as a clinical model.

METHODS

Sixty-one eyes in 61 patients were imaged by the Oxymap T1 Retinal Oximeter at baseline and 25 eyes were also examined 3 weeks after cataract surgery. Image quality (0-10 on a continuous scale) was compared with standardized AREDS cataract grading and Pentacam lens densitometry. Associations with retinal oximetry measurements and visual acuity were examined.

RESULTS

Image quality correlated with total, nuclear and posterior subcapsular cataract grades (ANOVA, p < 0.05), tended to be associated with lens densitometry and it improved from 4.3 ± 1.4 to 5.7 ± 1.0 (p < 0.05) after cataract surgery. Very low image quality, below 3, led to vessel detection failure in retinal oximetry images. Higher image qualities were linearly associated with higher measured retinal oxygen saturations (r = 0.52 in arteries and r = 0.46 in veins; p < 0.001).

CONCLUSION

Retinal oximetry image quality deteriorated with increasing cataract density and improved after cataract surgery, supporting its use as a measure of optical clarity. The numerical quality indicator demonstrated a threshold below which images of poor optical quality should be discarded. Image quality affects the estimates of retinal oximetry parameters and should therefore be included in future analyses.

Для цитирования: Klefter Oliver Niels, Erichsen Jesper Høiberg, Hansen Mathias Møller, Holm Lars Morten, Hardarson Sveinn Hakon, Stefánsson Einar, Kessel Line. Evaluation of a retinal oximetry image quality indicator in patients with cataract.. Acta ophthalmologica. 2023. DOI: 10.1111/aos.15747.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.