DOI 10.1007/s10384-023-01001-x
Extensive misalignment of plate-haptic rotationally asymmetric multifocal toric intraocular lens.
Аннотация
PURPOSE: Plate-haptic rotationally asymmetric multifocal toric intraocular lenses (IOL) (Lentis Comfort Toric) occasionally rotate extensively after surgery. We conducted the current study to investigate the incidence of extensive misalignment of this IOL and its association with clinical parameters.
STUDY
DESIGN
Retrospective case series.
SUBJECTS AND
METHODS
Data were collected from patients who had undergone phacoemulsification and implantation of a plate-haptic multifocal toric IOL.
RESULTS
Among 332 eyes, extensive misalignment of toric IOLs ≥ 45º occurred in 3.3% (11 eyes). The amount of misalignment in eyes with extensive misalignment was 81.6 ± 22.9º, while in those without extensive misalignment, it was 3.0 ± 2.7º. The eyes with extensive misalignment showed significantly greater axial length (p < 0.001), larger corneal diameter (p = 0.034), and flatter corneas (p = 0.044) than those without extensive misalignment. Repositioning surgery to correct toric IOL misorientation was conducted in 9 eyes between 7 and 28 days after cataract surgery. In 2 eyes, repositioning surgery was carried out twice.
CONCLUSIONS
In majority of cases plate-haptic multifocal toric IOLs showed satisfactory rotational stability, but extensive misalignment ≥ 45º occurred in 3.3% of cases.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.