Extensive misalignment of plate-haptic rotationally asymmetric multifocal toric intraocular lens.

Oshika Tetsuro, Nakano Shinichiro, Inamura Mikio, Ikai Nakako, Kato Yuji, Izumi Iori, Shimokawabe Kazuhito

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Plate-haptic rotationally asymmetric multifocal toric intraocular lenses (IOL) (Lentis Comfort Toric) occasionally rotate extensively after surgery. We conducted the current study to investigate the incidence of extensive misalignment of this IOL and its association with clinical parameters.

STUDY

DESIGN

Retrospective case series.

SUBJECTS AND

METHODS

Data were collected from patients who had undergone phacoemulsification and implantation of a plate-haptic multifocal toric IOL.

RESULTS

Among 332 eyes, extensive misalignment of toric IOLs ≥ 45º occurred in 3.3% (11 eyes). The amount of misalignment in eyes with extensive misalignment was 81.6 ± 22.9º, while in those without extensive misalignment, it was 3.0 ± 2.7º. The eyes with extensive misalignment showed significantly greater axial length (p < 0.001), larger corneal diameter (p = 0.034), and flatter corneas (p = 0.044) than those without extensive misalignment. Repositioning surgery to correct toric IOL misorientation was conducted in 9 eyes between 7 and 28 days after cataract surgery. In 2 eyes, repositioning surgery was carried out twice.

CONCLUSIONS

In majority of cases plate-haptic multifocal toric IOLs showed satisfactory rotational stability, but extensive misalignment ≥ 45º occurred in 3.3% of cases.

Для цитирования: Oshika Tetsuro, Nakano Shinichiro, Inamura Mikio, Ikai Nakako, Kato Yuji, Izumi Iori, Shimokawabe Kazuhito. Extensive misalignment of plate-haptic rotationally asymmetric multifocal toric intraocular lens.. Japanese journal of ophthalmology. 2023. DOI: 10.1007/s10384-023-01001-x.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.