DOI 10.3928/1081597x-20220701-01
Through-Focus Response of Extended Depth of Focus Intraocular Lenses.
Аннотация
PURPOSE: To investigate extended depth of focus intraocular lenses (EDOF IOLs), intended to extend the depth of field after cataract surgery in a comparative study at the optical bench.
METHODS
An optical bench with either green or white light was used for this study. The artificial cornea used exhibited a spherical aberration of 0.2 µm. Examinations of the following lenses with a pupil opening of 3 and 4.5 mm were carried out: AcrySof IQ Vivity (Alcon Laboratories, Inc), Isopure (PhysIOL), Tecnis Eyhance (Johnson & Johnson), Vivinex Impress (Hoya Surgical Optics), and xact (Santen).
RESULTS
Using green light and a pupil aperture of 3 mm, the AcrySof IQ Vivity showed the highest light energy for the intermediate area, whereas the Isopure and Vivinex Impress provided the highest energy for distance vision. Under the same examination conditions with a pupil opening of 4.5 mm, all lenses showed a low light distribution for the intermediate range. Regarding light distribution for distance, the Tecnis Eyhance had the highest light intensity. Using white light, the curves became much wider and more similar to each other.
CONCLUSIONS
The five EDOF lenses investigated differ mainly by their different weighting of energy between the far and intermediate ranges. [J Refract Surg. 2022;38(8):497-501.].
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.