DOI 10.1097/j.jcrs.0000000000000964
Advantageous effect of pupil dilation on the quality of optical biometry axial length measurement in individuals with dense cataract.
Аннотация
PURPOSE: To evaluate the effect of pharmacologic pupil dilation on the quality of axial length (AL) measurement in patients with dense cataracts and previous low-quality AL measurements performed without dilation.
SETTING
Shaare-Zedek Medical Center, Jerusalem, Israel.
DESIGN
Retrospective case series.
METHODS
All participants underwent swept-source optical coherence tomographic (SS-OCT) biometry. Those with low-quality AL measurements due to dense cataract underwent an additional SS-OCT biometric evaluation after pupil dilation, and the AL measurement of the same eye was recorded.
RESULTS
2076 patients underwent SS-OCT during the study period, and 177 (8.52%) had low-quality AL measurements because of dense cataract. 79 (44.63%, mean age 72.53 ± 13.27 years, 43 females) of those 177 patients underwent repeat SS-OCT biometric imaging after pharmacologic pupil dilation, and formed the study group. After pupil dilation, high-quality AL measurements were successfully obtained in 60 (75.95%) of the 79 eyes. The mean SD of the AL measurements decreased significantly ( P < .001), and the number of successful OCT scans per measurement increased significantly ( P < .001) after pupil dilation. The mean difference of the AL before and after pupil dilation was 0.03 ± 0.07 mm ( P < .001).
CONCLUSIONS
Pharmacologic pupil dilation improved the quality of SS-OCT biometrically measured AL in patients with low-quality AL measurement due to dense cataract. These results could potentially improve postoperative refractive outcomes after cataract surgery and reduce the need of additional AL measurements by more complicated alternative means in this group.