MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/lbr.0000000000000859

Медицинская библиотека · Journal of bronchology & interventional pulmonology · 2022

Complications of Endobronchial Valve Placement for Bronchoscopic Lung Volume Reduction: Insights From the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE).

Low See-Wei, Swanson Karen L, Lee Justin Z, Tan Min-Choon, Cartin-Ceba Rodrigo, Sakata Kenneth K, Maldonado Fabien

СтатьяПеревод пока не подготовлен

Аннотация

BACKGROUND

Patients with advanced emphysema experience breathlessness due to impaired respiratory mechanics and diaphragm dysfunction. Bronchoscopic lung volume reduction (BLVR) is a minimally invasive bronchoscopic procedure done to reduce hyperinflation and air trapping, promoting atelectasis in the targeted lobe and allowing improved respiratory mechanics. Real-world data on safety and complications outside of clinical trials of BLVR are limited.

METHODS

We queried the US Food and Drug Administrations (FDA) Manufacturers and User Device Experience database from May 2019 to June 2020 for reports involving BLVR with endobronchial valve (EBV) placement. Events were reviewed for data analysis.

RESULTS

We identified 124 cases of complications during BLVR with EBV implantation. The most-reported complication was pneumothorax (110/124, 89%), all of which required chest tube placement. A total of 54 of these cases (54/110, 49%) were complicated by persistent air leak requiring additional interventions. Repeat bronchoscopy was needed to remove the valves in 28 patients, 12 were discharged with a Heimlich valve, and 10 had an additional pleural catheter placed. The other complications of BLVR with EBV placement included respiratory failure (6/124, 5%), pneumonia (4/124, 3%), hemoptysis (2/124, 1.6%), valve migration (1/124, 1%), and pleural effusion (1/124, 1%). A total of 14 deaths were reported during that year.

CONCLUSION

Pneumothorax is the most-reported complication for BLVR with EBV placement, and in 65% of cases, pneumothorax is managed without removing valves. Importantly, 14 deaths were reported during that timeframe. Further studies are needed to estimate the true magnitude of the complications associated with BLVR.

Для цитирования: Low See-Wei, Swanson Karen L, Lee Justin Z, Tan Min-Choon, Cartin-Ceba Rodrigo, Sakata Kenneth K, Maldonado Fabien. Complications of Endobronchial Valve Placement for Bronchoscopic Lung Volume Reduction: Insights From the Food and Drug Administration Manufacturer and User Facility Device Experience (MAUDE).. Journal of bronchology & interventional pulmonology. 2022. DOI: 10.1097/lbr.0000000000000859.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.