BACKGROUND
Patients with advanced emphysema experience breathlessness due to impaired respiratory mechanics and diaphragm dysfunction. Bronchoscopic lung volume reduction (BLVR) is a minimally invasive bronchoscopic procedure done to reduce hyperinflation and air trapping, promoting atelectasis in the targeted lobe and allowing improved respiratory mechanics. Real-world data on safety and complications outside of clinical trials of BLVR are limited.
METHODS
We queried the US Food and Drug Administrations (FDA) Manufacturers and User Device Experience database from May 2019 to June 2020 for reports involving BLVR with endobronchial valve (EBV) placement. Events were reviewed for data analysis.
RESULTS
We identified 124 cases of complications during BLVR with EBV implantation. The most-reported complication was pneumothorax (110/124, 89%), all of which required chest tube placement. A total of 54 of these cases (54/110, 49%) were complicated by persistent air leak requiring additional interventions. Repeat bronchoscopy was needed to remove the valves in 28 patients, 12 were discharged with a Heimlich valve, and 10 had an additional pleural catheter placed. The other complications of BLVR with EBV placement included respiratory failure (6/124, 5%), pneumonia (4/124, 3%), hemoptysis (2/124, 1.6%), valve migration (1/124, 1%), and pleural effusion (1/124, 1%). A total of 14 deaths were reported during that year.
CONCLUSION
Pneumothorax is the most-reported complication for BLVR with EBV placement, and in 65% of cases, pneumothorax is managed without removing valves. Importantly, 14 deaths were reported during that timeframe. Further studies are needed to estimate the true magnitude of the complications associated with BLVR.