DOI 10.1097/j.jcrs.0000000000001968

Медицинская библиотека · Journal of cataract and refractive surgery · 2026

Evaluation of Optimised Implantable Collamer Lens Sizing using an Optical Low-Coherence Reflectometry Biometry Device at 12 months follow up.

Jain Nikhil, Laginaf Masara, Angunawela Romesh, Mearza Ali, Sherlock Johnathan, Barsam Allon

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: Implantable Collamer Lens (ICL) surgery can treat ametropia when laser vision correction is not suitable. Lens sizing is important to prevent insufficient or excessive vaulting. The Online Calculator and Ordering System (OCOS) requires an 'optimised' value for white-to-white (WTW) values when using the IOLMASTER or LENSTAR 900 (LS900) for biometry. We assessed if a more specific WTW subtraction for the LS900 could improve sizing outcomes.

SETTINGS: Single-centre London private practice setting.

DESIGN

A retrospective observational cohort study of consecutive patients undergoing ICL surgery within a 4-year period.

METHODS

Pre-operative measurements were taken using the LS900 and lenses were sized using the OCOS calculator with a WTW subtraction of 0.4mm. Vault height was measured at 2 weeks, 3 months and 12 months and complications recorded. Simulations were run using a more tailor-made subtraction of 0.25mm for the LS900.

RESULTS

199 eyes from 102 patients were analysed. 'Ideal' sizing occurred in 70% of ICLs at 3 months and dropped to 67% at 12 months. The largest proportion of incorrectly sized lenses were oversized at 3 months but undersized at 12 months. Simulations using an alternative WTW subtraction of 0.25mm suggest a redistribution of vault heights away from undersizing at 12 months when LS900 biometry is used.

CONCLUSION

Oversizing of ICLs using the OCOS calculator may be due to measurement discrepancies and early post-operative measurements of vault height. An alternative WTW subtraction of 0.25mm may reduce exposure to low vaults, which are a known risk factor for cataract formation. Real-world evaluation of this subtraction would be needed to confirm an improvement in accuracy.

Для цитирования: Jain Nikhil, Laginaf Masara, Angunawela Romesh, Mearza Ali, Sherlock Johnathan, Barsam Allon. Evaluation of Optimised Implantable Collamer Lens Sizing using an Optical Low-Coherence Reflectometry Biometry Device at 12 months follow up.. Journal of cataract and refractive surgery. 2026. DOI: 10.1097/j.jcrs.0000000000001968.
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.