MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.ajo.2026.04.014

Медицинская библиотека · American journal of ophthalmology · 2026

Intraindividual Comparison of Two Refractive Extended-Depth-of-Focus Intraocular Lens Designs.

Zeilinger Johannes, Palkovits Stefan, Ruiss Manuel, Pilwachs Caroline, Necpal Philipp, Elahi Ava, Findl Oliver

СтатьяСравнительное исследованиеРандомизированное исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: To compare intraindividual differences in visual performance of two refractive extended-depth-of-focus (EDF) intraocular lenses (IOLs) targeted for emmetropia.

DESIGN

Prospective, randomized, patient- and examiner-masked intraindividual comparative study.

METHODS

A total of 40 patients (80 eyes) with bilateral age-related cataract underwent phacoemulsification with implantation of refractive EDF IOL in each eye: a wavefront-shaped refractive EDF IOL (DEN00V, Johnson & Johnson Vision) in one eye and a sector-segmented EDF IOL (AN6V, Teleon) in the fellow eye. After 3 months, monocular distance- corrected and uncorrected distance (CDVA/UDVA) intermediate (DIVA/UIVA) and near (DNVA/UNVA) visual acuity, refractive outcomes, defocus curves, contrast sensitivity, halometry, and subjective visual quality using questionnaires were compared.

RESULTS

In 37 patients (74 eyes), mean UDVA at 4 m, DIVA at 66 cm and UDNVA at 40 cm were 0.05 ± 0.10, 0.17 ± 0.11, and 0.31 ± 0.11 logMAR for the DEN00V and 0.06 ± 0.10, 0.15 ± 0.10, and 0.32 ± 0.11 logMAR for the AN6V, respectively, with no significant differences. A statistically significant difference was observed for the primary endpoint, UIVA, with 0.18 ± 0.13 logMAR for the DEN00V and 0.12 ± 0.09 logMAR for the AN6V. Nominal differences were observed for CDVA and UNVA with 0.01 ± 0.07 and 0.32 ± 0.12 logMAR for the DEN00V and -0.02 ± 0.07 (P = .01) and 0.25 ± 0.10 (P = .01) logMAR for the AN6V. Further nominal differences could be shown for contrast sensitivity under photopic and mesopic conditions. However, these differences did not remain statistically significant after adjustment for multiple comparisons.

CONCLUSIONS

Both IOLs demonstrated an extended range of functional vision, preserved contrast sensitivity, and minimal photopic phenomena. A statistically significant difference was observed for the primary endpoint, UIVA, favoring the AN6V.

Для цитирования: Zeilinger Johannes, Palkovits Stefan, Ruiss Manuel, Pilwachs Caroline, Necpal Philipp, Elahi Ava, Findl Oliver. Intraindividual Comparison of Two Refractive Extended-Depth-of-Focus Intraocular Lens Designs.. American journal of ophthalmology. 2026. DOI: 10.1016/j.ajo.2026.04.014.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.