DOI 10.3390/s24155027

Медицинская библиотека · Sensors (Basel, Switzerland) · 2024

Continuous Temperature Telemonitoring of Patients with COVID-19 and Other Infectious Diseases Treated in Hospital-at-Home: Viture® System Validation.

Sesma-Sánchez Laura, Ruiz-Castellano María, Romero-Roldán Ainara, Álvarez-García Laura, Morrás-Gómez Marta, Tabar-Liberal Idoia, Pulido-Fontes Marta, Salmón-García Belén

СтатьяВалидационное исследованиеПеревод пока не подготовлен

Аннотация

Body temperature must be monitored in patients receiving Hospital-at-Home (HaH) care for COVID-19 and other infectious diseases. Continuous temperature telemonitoring (CTT) detects fever and patient deterioration early, facilitating decision-making. We performed a validation clinical study assessing the safety, comfort, and impact on healthcare practice of Viture®, a CTT system, compared with a standard digital axillary thermometer in 208 patients with COVID-19 and other infectious diseases treated in HaH at the Navarra University Hospital (HUN). Overall, 3258 pairs of measurements showed a clinical bias of -0.02 °C with limits of agreement of -0.96/+0.92 °C, a 95% acceptance rate, and a mean absolute deviation of 0.36 (SD 0.30) °C. Viture® detected 3 times more febrile episodes and revealed fever in 50% more patients compared with spot measurements. Febrile episodes were detected 7.23 h (mean) earlier and modified the diagnostic and/or therapeutic approach in 43.2% of patients. Viture® was validated for use in a clinical setting and was more effective in detecting febrile episodes than conventional methods.

Для цитирования: Sesma-Sánchez Laura, Ruiz-Castellano María, Romero-Roldán Ainara, Álvarez-García Laura, Morrás-Gómez Marta, Tabar-Liberal Idoia, Pulido-Fontes Marta, Salmón-García Belén. Continuous Temperature Telemonitoring of Patients with COVID-19 and Other Infectious Diseases Treated in Hospital-at-Home: Viture® System Validation.. Sensors (Basel, Switzerland). 2024. DOI: 10.3390/s24155027.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.