Volumetric Measurement of Optic Nerve Head Drusen Using Swept-Source Optical Coherence Tomography.

Tsikata Edem, Verticchio Vercellin Alice C, Falkenstein Iryna, Poon Linda Yi-Chieh, Brauner Stacey, Khoueir Ziad, Miller John B, Chen Teresa C

СтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To describe new software tools for quantifying optic nerve head drusen volume using 3-dimensional (3D) swept-source optical coherence tomography (SS-OCT) volumetric scans.

MATERIALS AND

METHODS

SS-OCT was used to acquire raster volume scans of 8 eyes of 4 patients with bilateral optic nerve head drusen. The scans were manually segmented by 3 graders to identify the drusen borders, and thereafter total drusen volumes were calculated. Linear regression was performed to study the relationships between drusen volume, retinal nerve fiber layer thickness, and Humphrey visual field mean deviation.

RESULTS

In the 8 study eyes, drusen volumes ranged between 0.24 to 1.05 mm. Visual field mean deviation decreased by ∼20 dB per cubic millimeter increase in drusen volume, and the coefficient of correlation of the linear regression was 0.92. In this small patient series, visual field defects were detected when drusen volume was larger than about 0.2 mm.

CONCLUSIONS

Software tools have been developed to quantify the size of OHND using SS-OCT volume scans.

Для цитирования: Tsikata Edem, Verticchio Vercellin Alice C, Falkenstein Iryna, Poon Linda Yi-Chieh, Brauner Stacey, Khoueir Ziad, Miller John B, Chen Teresa C. Volumetric Measurement of Optic Nerve Head Drusen Using Swept-Source Optical Coherence Tomography.. Journal of glaucoma. 2017. DOI: 10.1097/ijg.0000000000000707.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.