OBJECTIVES

The aim of this study was to analyse long-term results of aortic root replacement with the Shelhigh® NR-2000C conduit.

METHODS

From January 2001 to October 2005, 63 patients with a median age of 62 years underwent aortic root replacement with a Shelhigh® conduit. Aneurysm (27%), aortic valve endocarditis (30%) and acute type A aortic dissection (33%) were the predominant indications for the surgery. Fifty-four patients (86%) were entered in the follow-up study with the end-points of death, aortic root-related reoperation and endocarditis.

RESULTS

The overall 30-day mortality rate was 8% (n = 5). The median follow-up was 9.5 years (range 0-14.2 years). In total, 13 (24%) deaths occurred during follow-up; of these, 4 were directly conduit-related. The overall estimated survival rates at 1, 5 and 10 years were 85 ± 5, 79 ± 6 and 71 ± 7%, respectively. Reoperation was necessary in 10 (19%) patients due to endocarditis (n = 5), aortic stenosis (n = 3), pseudoaneurysm due to detachment of the right coronary artery (n = 1) and detachment of the non-coronary leaflet (n = 1). The overall estimated rate of freedom from aortic root-related reoperation at 1, 5 and 10 years was 83 ± 5, 79 ± 6 and 64 ± 7%, respectively. Endocarditis of the prosthesis was reported in 9 (17%) patients; of whom, 5 patients required reoperation and 4 were treated medically. In 1 patient with endocarditis, a stroke was reported due to a thromboembolic event.

CONCLUSIONS

The first long-term follow-up after aortic root replacement with the Shelhigh® BioConduit revealed a relatively high rate of death and very high rate of reoperations due to endocarditis, aorto-ventricular disconnection and structural valve failure. This may be potentially connected to the nature of the implanted valved conduit.

Для цитирования: Sahin Ayhan, Müggler Oliver, Sromicki Juri, Caliskan Etem, Reser Diana, Emmert Maximilian Y, Alkadhi Hatem, Maisano Francesco, Falk Volkmar, Holubec Tomas. Long-term follow-up after aortic root replacement with the Shelhigh® biological valved conduit: a word of caution!. European journal of cardio-thoracic surgery : official journal of the European Association for Cardio-thoracic Surgery. 2016. DOI: 10.1093/ejcts/ezw167.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.