MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/ijg.0000000000002013

Медицинская библиотека · Journal of glaucoma · 2022

Clinical Outcomes of Ahmed ClearPath Implantation in Glaucomatous Eyes: A Novel Valveless Glaucoma Drainage Device.

Grover Davinder S, Kahook Malik Y, Seibold Leonard K, Singh Inder Paul, Ansari Husam, Butler Michelle R, Smith Oluwatosin U, Sawhney Gagan K, Van Tassel Sarah H, Dorairaj Syril

СтатьяМногоцентровое исследованиеПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PRCIS: Designed with novel features to facilitate implantation and improve safety, the Ahmed ClearPath (ACP) glaucoma drainage device (GDD) provided intraocular pressure (IOP) reduction comparable with other GDDs in eyes with refractory glaucoma in a multicenter retrospective study.

PURPOSE: To present clinical outcomes with a novel valveless GDD (ACP, New World Medical). The 250 and 350 mm2 models feature a contoured plate for optimal globe apposition, anteriorized suture points to facilitate suturing to the globe, and a prethreaded 4-0 polypropylene ripcord suture.

METHODS

This was a multicenter retrospective analysis of eyes with medically and/or surgically uncontrolled glaucoma implanted with the 250 or 350 mm2 ACP either as a standalone procedure or in combination with other procedures. Pre-, intra-, and postoperative data through 6 months were collected.

RESULTS

A total of 104 eyes (100 subjects) received the ACP by 10 US surgeons, 63.5% of which had primary open-angle glaucoma and 62.5% had severe glaucoma. Mean baseline IOP was 26.3 (9.0) mm Hg and mean medication use was 3.9 (1.3). Through 6 months' follow-up, mean IOP ranged from 13.6 to 16.7 mm Hg and mean medication use from 0.9 to 1.9 medications (P<0.0001 at all timepoints for each outcome measure). At 6 months, mean IOP was 13.7 mm Hg (-13.0 mm Hg, 43.0%, P<0.0001) and mean medication use was 1.9 medications per eye (-2.1, 47.7%, P<0.0001). Common adverse events included anterior chamber inflammation (16.3%), hyphema (15.4%), and hypotony (6.7%).

CONCLUSION

The new ACP appears to be safe and efficacious as a standalone procedure or in combination with other procedures for uncontrolled glaucoma, and may be considered as a GDD option for patients in whom its unique design may facilitate the implantation process.

Для цитирования: Grover Davinder S, Kahook Malik Y, Seibold Leonard K, Singh Inder Paul, Ansari Husam, Butler Michelle R, Smith Oluwatosin U, Sawhney Gagan K, Van Tassel Sarah H, Dorairaj Syril. Clinical Outcomes of Ahmed ClearPath Implantation in Glaucomatous Eyes: A Novel Valveless Glaucoma Drainage Device.. Journal of glaucoma. 2022. DOI: 10.1097/ijg.0000000000002013.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.