MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1111/jce.12431

Медицинская библиотека · Journal of cardiovascular electrophysiology · 2014

Fluoroscopy of spontaneous breathing is more sensitive than phrenic nerve stimulation for detection of right phrenic nerve injury during cryoballoon ablation of atrial fibrillation.

Linhart Markus, Nielson Annika, Andrié René P, Mittmann-Braun Erica L, Stöckigt Florian, Kreuz Jens, Nickenig Georg, Schrickel Jan W, Lickfett Lars M

СтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

Right phrenic nerve palsy (PNP) is a typical complication of cryoballoon ablation of the right-sided pulmonary veins (PVs). Phrenic nerve function can be monitored by palpating the abdomen during phrenic nerve pacing from the superior vena cava (SVC pacing) or by fluoroscopy of spontaneous breathing. We sought to compare the sensitivity of these 2 techniques during cryoballoon ablation for detection of PNP.

METHODS AND

RESULTS

A total of 133 patients undergoing cryoballoon ablation were monitored with both SVC pacing and fluoroscopy of spontaneous breathing during ablation of the right superior PV. PNP occurred in 27/133 patients (20.0%). Most patients (89%) had spontaneous recovery of phrenic nerve function at the end of the procedure or on the following day. Three patients were discharged with persistent PNP. All PNP were detected first by fluoroscopic observation of diaphragm movement during spontaneous breathing, while diaphragm could still be stimulated by SVC pacing. In patients with no recovery until discharge, PNP occurred at a significantly earlier time (86 ± 34 seconds vs. 296 ± 159 seconds, P < 0.001). No recovery occurred in 2/4 patients who were ablated with a 23 mm cryoballoon as opposed to 1/23 patients with a 28 mm cryoballoon (P = 0.049).

CONCLUSION

Fluoroscopic assessment of diaphragm movement during spontaneous breathing is more sensitive for detection PNP as compared to SVC pacing. PNP as assessed by fluoroscopy is frequent (20.0%) and carries a high rate of recovery (89%) until discharge. Early onset of PNP and use of 23 mm cryoballoon are associated with PNP persisting beyond hospital discharge.

Для цитирования: Linhart Markus, Nielson Annika, Andrié René P, Mittmann-Braun Erica L, Stöckigt Florian, Kreuz Jens, Nickenig Georg, Schrickel Jan W, Lickfett Lars M. Fluoroscopy of spontaneous breathing is more sensitive than phrenic nerve stimulation for detection of right phrenic nerve injury during cryoballoon ablation of atrial fibrillation.. Journal of cardiovascular electrophysiology. 2014. DOI: 10.1111/jce.12431.