A Case Report of Complete Blockage of a Baerveldt Glaucoma Implant Following Insertion of a 3-0 Supramid Suture.

Lee Richard M H, Bouremel Yann, Eames Ian, Brocchini Steve, Khaw Peng Tee

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

PURPOSE: The aim of this study was to present a case of a Baerveldt glaucoma implant lumen being completely occluded with a 3-0 Supramid stent suture.

PATIENT AND

METHODS

The patient underwent Baerveldt glaucoma implant surgery with placement of an intraluminal 3-0 Supramid stent suture that acts to restrict flow across the device and reduce the risk of postoperative hypotony. Following suturing of the implant to the sclera, the device was flow tested. No flow was observed through the device tube and a significant ballooning of the tube diameter occurred with increased pressure on the device. The device was explanted from the eye and replaced with a different implant without further postoperative complication. The explanted device was assessed using custom microfluidic equipment in an in vitro environment.

RESULTS

This phenomenon occurred despite using several different batches of the 3-0 Supramid stent suture and the device had to be removed and replaced with another device without complication. In vitro microfluidic assessment of the device demonstrated no flow across the device tube despite over 150 mm Hg of pressure being exerted on the device.

CONCLUSIONS

We hypothesize that the blockage occurred at the junction between the device tube and plate and that the ballooning phenomenon observed was due to a defect in the tube wall. This case highlights the importance of flow testing all glaucoma drainage devices before insertion given the variation in manufacturing conditions to avoid the risk of intraoperative complications.

Для цитирования: Lee Richard M H, Bouremel Yann, Eames Ian, Brocchini Steve, Khaw Peng Tee. A Case Report of Complete Blockage of a Baerveldt Glaucoma Implant Following Insertion of a 3-0 Supramid Suture.. Journal of glaucoma. 2019. DOI: 10.1097/ijg.0000000000001198.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.