MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euu339

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2015

Validation of the mapping accuracy of a novel non-invasive epicardial and endocardial electrophysiology system.

Revishvili Amiran S, Wissner Erik, Lebedev Dmitry S, Lemes Christine, Deiss Sebastian, Metzner Andreaas, Kalinin Vitaly V, Sopov Oleg V, Labartkava Eugeny Z, Kalinin Alexander V, Chmelevsky Michail, Zubarev Stephan V, Chaykovskaya Maria K, Tsiklauri Mikhail G, Kuck Karl-Heinz

Клиническое исследованиеСтатьяПоддержка исследования: вне правительства СШАВалидационное исследованиеПеревод пока не подготовлен

Аннотация

AIMS: Use of a non-invasive electrocardiographic mapping system may aid in rapid diagnosis of atrial or ventricular arrhythmias or the detection of ventricular dyssynchrony. The aim of the present study was to validate the mapping accuracy of a novel non-invasive epi- and endocardial electrophysiology system (NEEES).

METHODS AND

RESULTS

Patients underwent pre-procedural computed tomography or magnetic resonance imaging of the heart and torso. Radiographic data were merged with the data obtained from the NEEES during pacing from implanted pacemaker leads or pacing from endocardial sites using an electroanatomical mapping system (CARTO 3, Biosense Webster). The earliest activation as denoted on the NEEES three-dimensional heart model was compared with the true anatomic location of the tip of the pacemaker lead or the annotated pacing site on the CARTO 3 map. Twenty-nine patients [mean age: 62 ± 11 years, 6/29 (11%) female, 21/29 (72%) with ischaemic cardiomyopathy] were enrolled into the pacemaker verification group. The mean distance from the non-invasively predicted pacing site to the anatomic reference site was 10.8 ± 5.4 mm for the right atrium, 7.7 ± 5.8 mm for the right ventricle, and 7.9 ± 5.7 mm for the left ventricle activated via the coronary sinus lead. Five patients [mean age 65 ± 4 years, 2 (33%) females] underwent CARTO 3 verification study. The mean distance between non-invasively reconstructed pacing site and the reference pacing site was 7.4 ± 2.7 mm for the right atrium, 6.9 ± 2.3 mm for the left atrium, 6.5 ± 2.1 mm for the right ventricle, and 6.4 ± 2.2 for the left ventricle, respectively.

CONCLUSION

The novel NEEES was able to correctly identify the site of pacing from various endo- and epicardial sites with high accuracy.

Для цитирования: Revishvili Amiran S, Wissner Erik, Lebedev Dmitry S, Lemes Christine, Deiss Sebastian, Metzner Andreaas, Kalinin Vitaly V, Sopov Oleg V, Labartkava Eugeny Z, Kalinin Alexander V, Chmelevsky Michail, Zubarev Stephan V, Chaykovskaya Maria K, Tsiklauri Mikhail G, Kuck Karl-Heinz. Validation of the mapping accuracy of a novel non-invasive epicardial and endocardial electrophysiology system.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2015. DOI: 10.1093/europace/euu339.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.