Haemodynamic vector personalization of a quadripolar left ventricular lead used for cardiac resynchronization therapy: use of surface electrocardiogram and interventricular time delays.

Trolese Luca, Biermann Juergen, Hartmann Maximilian, Schluermann Fabienne, Faber Thomas S, Bode Christoph, Asbach Stefan

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

AIMS: The choice of left ventricular pacing configurations (LVPCs) of quadripolar leads used for cardiac resynchronization therapy (CRT) affects haemodynamic response and thus may be a tool for device optimization. The value of surface electrocardiograms and interventricular time delays (IVDs) for optimization is unknown.

METHODS AND

RESULTS

Sixteen patients implanted with a CRT device with a quadripolar LV lead underwent invasive testing of LV dP/dt. QRS durations at baseline (bl) and during biventricular pacing (biv) were measured using different LVPCs (total of 141 LVPCs; 8.8 per patient). Variations in QRS duration during biv were calculated for each patient (ΔQRS) and, when compared with intrinsic QRS duration, for all LVPCs (ΔQRSLVPC). Interventricular time delays between the poles of the LV lead were obtained from intracardiac electrograms. ΔIVD was calculated as IVDmax - IVDmin. Parameters were correlated with LV dP/dt. ΔQRS and ΔQRSLVPC both significantly correlated with LV dP/dt (P < 0.01). Correlation was found for patients with ischaemic (P < 0.001) and non-ischaemic cardiomyopathy (P < 0.05), and for patients with bl QRS duration >168 ms (P < 0.001), but not <168 ms (P = ns). The LVPC with shortest QRS duration also yielded maximal LV dP/dt in 6 of 16 patients (37.5%), and was equal or better in LV dP/dt in 12 of 16 patients (75%). ΔIVD neither correlated with ΔQRS nor ΔLV dP/dt.

CONCLUSION

ΔQRS predicts the maximal value of vector personalization in the individual. Reductions in QRS width, but not IVDs, correlate with acute haemodynamic response. Intraindividually, in 75% of patients, the LVPC with the shortest QRS duration gives equal or superior haemodynamic results when compared with the LVPC with longest QRS duration.

Для цитирования: Trolese Luca, Biermann Juergen, Hartmann Maximilian, Schluermann Fabienne, Faber Thomas S, Bode Christoph, Asbach Stefan. Haemodynamic vector personalization of a quadripolar left ventricular lead used for cardiac resynchronization therapy: use of surface electrocardiogram and interventricular time delays.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2014. DOI: 10.1093/europace/euu136.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.