Sleeping with the fishes: electromagnetic interference causing an inappropriate implantable cardioverter defibrillator shock.

Knight Hui Min, Cakebread Holly Elizabeth, Gajendragadkar Parag Ravindra, Duehmke Rudolf Martin

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 60-year-old man with a cardiac defibrillator implanted due to previous ventricular fibrillation arrest and ischaemic cardiomyopathy received a shock while cleaning his fish pond. At the time, his immersed arm was close to a submersed water pump, but the patient was asymptomatic. As a result of the shock he lost consciousness, but collapsed backwards, away from the pond. Interrogation of the device revealed a high-frequency artefact that was sensed by the device and triggered a shock. Device parameters were otherwise normal. Subsequently, the submersed water pump was found to be the source of an external alternating current leak and was identified as the likely cause of the inappropriate shock due to electromagnetic interference (EMI). Awareness of potential sources of EMI along with evaluation of data with a detailed clinical history is warranted in all cases.

Для цитирования: Knight Hui Min, Cakebread Holly Elizabeth, Gajendragadkar Parag Ravindra, Duehmke Rudolf Martin. Sleeping with the fishes: electromagnetic interference causing an inappropriate implantable cardioverter defibrillator shock.. BMJ case reports. 2014. DOI: 10.1136/bcr-2013-203462.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.