MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/eur143

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2011

Effective prediction of response to cardiac resynchronization therapy using a novel program of gated myocardial perfusion single photon emission computed tomography.

Nakamura Kentaro, Takami Mitsuaki, Shimabukuro Michio, Maesato Akira, Chinen Ichiro, Ishigaki Sugako, Higa Satoshi, Keida Takehiko, Masuzaki Hiroaki

Оценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

AIMS: The limitation of QRS duration as a surrogate measure for left ventricular (LV) mechanical dyssynchrony (LVMD) in cardiac resynchronization therapy (CRT) patient selection encourages seeking alternatives to QRS duration. Exploring the potential of an analysis program of electrocardiographically gated myocardial perfusion single photon emission computed tomography (SPECT) (GMPS) for the estimation of LVMD to predict CRT response.

METHODS AND

RESULTS

Twenty-four patients undergoing CRT for advanced heart failure caused by non-ischaemic cardiomyopathy were studied. Gated myocardial perfusion single photon emission computed tomographies were performed in the setting of temporary CRT suspension after 1 week of CRT adoption. The GMPS data were computed with a novel program capable of segmental LV time-volume analysis. When a brain natriuretic peptide (BNP) value decreased >50% at 6-month follow-up, the patient was defined as a CRT responder. Receiver operating characteristic (ROC) curves for identification of responders were analysed for standard deviation of time to end systole (TES-SD) among 17 LV segments. Linear regression analyses demonstrated that an increase in percentage reduction in BNP level at 6-month follow-up was predicted by an increase in TES-SD (R(2) = 0.21, P = 0.023). The TES-SD in responders (n = 15, 62.5%) was higher than that in non-responders (100 ± 51 vs. 41 ± 17 ms, P = 0.0008). A cutoff value of TES-SD >49 ms predicted responders with 100% sensitivity and 78.8% specificity and the area under the ROC curve was 0.881 for TES-SD (P = 0.002).

CONCLUSIONS

The estimation of LVMD using this novel GMPS program could be an alternative or a complementary approach to QRS duration in CRT patient selection. This finding warrants further assessment of our approach in larger studies.

Для цитирования: Nakamura Kentaro, Takami Mitsuaki, Shimabukuro Michio, Maesato Akira, Chinen Ichiro, Ishigaki Sugako, Higa Satoshi, Keida Takehiko, Masuzaki Hiroaki. Effective prediction of response to cardiac resynchronization therapy using a novel program of gated myocardial perfusion single photon emission computed tomography.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2011. DOI: 10.1093/europace/eur143.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.