Relationship between left ventricular stimulation characteristics at implantation and echocardiographic response after 6 months of cardiac resynchronization therapy.

Deplagne Antoine, Ploux Sylvain, Ritter Philippe, Lafitte Stephane, Reant Patricia, Jais Pierre, Haissaguerre Michel, Clementy Jacques, Bordachar Pierre

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Although the electrical stimulation of an ischaemic tissue adversely affects the left ventricular (LV) systolic function, the optimal stimulation site in patients with non-ischaemic cardiomyopathy has not been systematically studied. We hypothesized that the local stimulation characteristics at the time of device implantation predict the response to cardiac resynchronization therapy (CRT).

METHODS AND

RESULTS

We measured the impedance, sensing, and capture threshold of a bipolar LV lead in 138 patients with non-ischaemic cardiomyopathy undergoing first implantation of CRT device for drug refractory heart failure. All patients underwent echocardiography at baseline and at 6 months post-implantation. An absence of favourable response to CRT was defined as <15% decrease in echocardiographic LV end-systolic volume (LVESV) at 6 months. Echocardiographic response to CRT was observed in 70% of patients. The LV lead measurements predicted neither the optimal stimulation site nor the response to CRT. Left ventricular capture threshold (1.50 ± 1.1 vs. 1.32 ± 0.8 V) and impedance (725 ± 287 vs. 720 ± 261 Ω) were similar between the responders and the non-responders. Independent of baseline LV ejection fraction or ESV, the LV R-wave amplitude at implantation was significantly higher (P = 0.0038) in responders (12.7 ± 5.2 mV) than in non-responders (9.7 ± 6.3 mV), with an area under the receiver operating characteristic curve of 0.7.

CONCLUSION

Response to CRT, as determined by decrease in LVESV at 6 months, was associated with significantly higher LV R-wave amplitude at the time of device implantation.

Для цитирования: Deplagne Antoine, Ploux Sylvain, Ritter Philippe, Lafitte Stephane, Reant Patricia, Jais Pierre, Haissaguerre Michel, Clementy Jacques, Bordachar Pierre. Relationship between left ventricular stimulation characteristics at implantation and echocardiographic response after 6 months of cardiac resynchronization therapy.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2010. DOI: 10.1093/europace/euq301.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.