Real-time three-dimensional speckle tracking echocardiography: a novel technique to quantify global left ventricular mechanical dyssynchrony.

Thebault Christophe, Donal Erwan, Bernard Anne, Moreau Olivier, Schnell Frederic, Mabo Philippe, Leclercq Christophe

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: left ventricular (LV) mechanical dyssynchrony (LVMD) has emerged as a therapeutic target using cardiac resynchronization therapy (CRT) in chronic heart failure patients. Current methods used to evaluate LVMD are technically challenging and do not assess all the components of LVMD simultaneously. We analysed real-time 3D speckle tracking (3DST) echocardiography as a novel method to assess LVMD.

METHODS AND

RESULTS

three-dimensional ST was performed in 60 unselected patients (71 ± 9 years old; 39% with ischaemic cardiomyopathy) who were referred to optimize and to control of a CRT device; implanted according to current guidelines (9 added to be excluded from the protocol). Two standardized conditions [right ventricular (RV) versus an optimized biventricular (BiV) pacing modality] were tested. These two pacing modalities lead to two distinct electrical activation patterns. We sought to test the capability of 3DST to distinguish these two patterns. The LV ejection fraction was 24 ± 9 in the RV mode and 29 ± 10% in BiV. By 3DST, we measured global end-systolic LV deformation and dyssynchrony (standard deviation of the time to peak/16 LV-segment). The 3D radial strain increased from 13.8 ± 5.7 in the RV to 15.9 ± 6.5% in theBiV mode, and the dyssynchrony index decreased from 15.1 ± 5.0 to 11.8 ± 4.1%. 3D longitudinal strain increased from -6.9 ± 2.8 in the RV to -7.8 ± 3.2% in the BiV mode, and the dyssynchrony index decreased from 14.2 ± 4.8 to 11.5 ± 5.0% (P < 0.01 for all). The 3D area strain (AS) increased from -15.4 ± 6 in the RV to -18.3 ± 7.0% in the BiV mode, and the dyssynchrony index decreased from 12.2 ± 5.1 to 9.5 ± 4.5% (P < 0.001 for all).

CONCLUSION

When image quality is optimal, 3DST might offer a new rapid method to quantify global LVMD in CRT candidates. In a comparison of the utility of various 3D strain measurements, the 3D AS appears to be closed to the ideal parameter that we are looking for.

Для цитирования: Thebault Christophe, Donal Erwan, Bernard Anne, Moreau Olivier, Schnell Frederic, Mabo Philippe, Leclercq Christophe. Real-time three-dimensional speckle tracking echocardiography: a novel technique to quantify global left ventricular mechanical dyssynchrony.. European journal of echocardiography : the journal of the Working Group on Echocardiography of the European Society of Cardiology. 2010. DOI: 10.1093/ejechocard/jeq095.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.