Impact of baseline impedance of pulmonary vein antrum on success of catheter ablation for paroxysmal atrial fibrillation guided by ablation index.

Sun Yuanjun, Xiao Xianjie, Yin Xiaomeng, Gao Lianjun, Yu Xiaohong, Zhang Rongfeng, Wang Zhongzhen, Dai Shiyu, Yang Yanzong, Xia Yunlong

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

Ablation index (AI) is an effective ablation quality marker. Impedance is also an important factor for lesion formation. The present study evaluated the influence of the baseline impedance in the effect of ablation for atrial fibrillation (AF) guided by AI.

METHODS

This was a retrospective study. 101 patients with paroxysmal AF (PAF) were enrolled. All patients underwent radiofrequency ablation guided by the same AI strategy. The ablation strategy was pulmonary vein (PV) isolation with non-PV triggers ablation. The baseline impedance of the ablation points was recorded. The patients were followed up every 3 months or so.

RESULTS

During a median follow-up of 12 (4-14) months, freedom from AF/atrial tachycardia recurrence were 82.2%. No difference existed in baseline characteristics between the success group and the recurrence group. The average baseline impedance was 124.3 ± 9.7 Ω. The baseline impedance of the ablation points in success group was lower compared to the recurrence group (122.9 ± 9.4 vs. 130.5 ± 8.8 Ω, P < 0.01). The ratio of impedance drop in the success group was higher than the recurrence group ([8.8 ± 1.4]% vs. [8.1 ± 1.2]%, P = 0.03). Multivariate analysis revealed that baseline impedance, PAF duration and AI were the independent predictors of AF recurrence. The cumulative free of recurrence rate of low-impedance group (≤ 124 Ω, n = 54) was higher than that of high-impedance group.

CONCLUSION

Baseline impedance correlates with clinical outcome of radiofrequency ablation for PAF guided by AI. Higher impedance in the same AI strategy may result in an ineffective lesion which probably causes recurrence.

Для цитирования: Sun Yuanjun, Xiao Xianjie, Yin Xiaomeng, Gao Lianjun, Yu Xiaohong, Zhang Rongfeng, Wang Zhongzhen, Dai Shiyu, Yang Yanzong, Xia Yunlong. Impact of baseline impedance of pulmonary vein antrum on success of catheter ablation for paroxysmal atrial fibrillation guided by ablation index.. BMC cardiovascular disorders. 2022. DOI: 10.1186/s12872-022-02530-y.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.