DOI 10.1007/s10840-020-00780-4

Медицинская библиотека · Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing · 2020

Long-term safety and effectiveness of paroxysmal atrial fibrillation ablation using a porous tip contact force-sensing catheter from the SMART SF trial.

Natale Andrea, Monir George, Patel Anshul M, Fishel Robert S, Marchlinski Francis E, Delaughter M Craig, Athill Charles A, Melby Daniel P, Gonzalez Mario D, Hariharan Ramesh, Gidney Brett, Tan Tiffany, Chinitz Larry A

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

PURPOSE: The prospective, multicenter SMART SF trial demonstrated the acute safety and effectiveness of the 56-hole porous tip irrigated contact force (CF) catheter for drug-refractory paroxysmal atrial fibrillation (PAF) ablation with a low primary adverse event rate (2.5%), leading to FDA approval of the catheter. Here, we are reporting the long-term effectiveness and safety results that have not yet been reported.

METHODS

Ablations were performed using the 56-hole porous tip irrigated CF catheter guided by the 3D mapping system stability module. The primary effectiveness endpoint was freedom from atrial tachyarrhythmia (including atrial fibrillation, atrial tachycardia, and/or atrial flutter), based on electrocardiographic data at 12 months. Atrial tachyarrhythmia recurrence occurring 3 months post procedure, acute procedural failures such as lack of entrance block confirmation of all PVs, and undergoing repeat procedure for atrial fibrillation in the evaluation period (91 to 365 days post the initial ablation procedure) were considered to be effectiveness failures.

RESULTS

Seventy-eight patients (age 64.8 ± 9.7 years; male 52.6%; Caucasian 96.2%) participated in the 12-month effectiveness evaluation. Mean follow-up time was 373.5 ± 45.4 days. The Kaplan-Meier estimate of freedom from 12-month atrial tachyarrhythmia was 74.9%. Two procedure-related pericardial effusion events were reported at 92 and 180 days post procedure. There were no pulmonary vein stenosis complications or deaths reported through the 12-month follow-up period.

CONCLUSIONS

The SMART SF 12-month follow-up evaluation corroborates the early safety and effectiveness success previously reported for PAF ablation with STSF.

Для цитирования: Natale Andrea, Monir George, Patel Anshul M, Fishel Robert S, Marchlinski Francis E, Delaughter M Craig, Athill Charles A, Melby Daniel P, Gonzalez Mario D, Hariharan Ramesh, Gidney Brett, Tan Tiffany, Chinitz Larry A. Long-term safety and effectiveness of paroxysmal atrial fibrillation ablation using a porous tip contact force-sensing catheter from the SMART SF trial.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2020. DOI: 10.1007/s10840-020-00780-4.