MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1093/europace/euae016

Медицинская библиотека · Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology · 2023

Pulsed electric field, cryoballoon, and radiofrequency for paroxysmal atrial fibrillation ablation: a propensity score-matched comparison.

Della Rocca Domenico G, Marcon Lorenzo, Magnocavallo Michele, Menè Roberto, Pannone Luigi, Mohanty Sanghamitra, Sousonis Vasileios, Sorgente Antonio, Almorad Alexandre, Bisignani Antonio, Głowniak Andrzej, Del Monte Alvise, Bala Gezim, Polselli Marco, Mouram Sahar, La Fazia Vincenzo Fazia, Ströker Erwin, Gianni Carola, Zeriouh Sarah, Bianchi Stefano, Sieira Juan, Combes Stephane, Sarkozy Andrea, Rossi Pietro, Boveda Serge, Natale Andrea, de Asmundis Carlo, Chierchia Gian-Battista, HRMC Investigators

СтатьяПеревод пока не подготовлен

Аннотация

AIMS: Pulsed field ablation (PFA) has emerged as a novel, non-thermal energy source to selectively ablate cardiac tissue. We describe a multicentre experience on pulmonary vein isolation (PVI) via the pentaspline Farapulse™ PFA system vs. thermal-based technologies in a propensity score-matched population of paroxysmal atrial fibrillation (PAF) patients.

METHODS AND

RESULTS

Propensity score matching was adopted to compare PVI-only ablation outcomes via the Farawave™ system (Group PFA), cryoballoon (Group CRYO), or focal radiofrequency (Group RF) (PFA:CRYO:RF ratio = 1:2:2). Among 1572 (mean age: 62.4 ± 11.3 years; 42.5% females) PAF patients undergoing first time PVI with either PFA (n = 174), CRYO (n = 655), or RF (n = 743), propensity score matching yielded 174 PFA, 348 CRYO, and 348 RF patients. First-pass isolation was achieved in 98.8% of pulmonary veins (PVs) with PFA, 81.5% with CRYO, and 73.1% with RF (P < 0.001). Procedural and dwell times were significantly shorter with PFA, whereas the availability of a 3D mapping system led to a significant reduction in X-ray exposure with RF. Overall complication rates were 3.4% (n = 6) with PFA, 8.6% (n = 30) with CRYO, and 5.5% (n = 19) with RF (P = 0.052). The 1-year Kaplan-Meier estimated freedom from any atrial tachyarrhythmia was 79.3% with PFA, 74.7% with CRYO, and 72.4% with RF (log-rank P-value: 0.24). Among 145 repeat ablation procedures, PV reconnection rate was 19.1% after PFA, 27.5% after CRYO, and 34.8% after RF (P = 0.01).

CONCLUSION

Pulsed field ablation contributed to significantly shorter procedural times. Follow-up data showed a similar arrhythmia freedom, although a higher rate of PV reconnection was documented in post-CRYO and post-RF redo procedures.

Для цитирования: Della Rocca Domenico G, Marcon Lorenzo, Magnocavallo Michele, Menè Roberto, Pannone Luigi, Mohanty Sanghamitra, Sousonis Vasileios, Sorgente Antonio, Almorad Alexandre, Bisignani Antonio, Głowniak Andrzej, Del Monte Alvise, Bala Gezim, Polselli Marco, Mouram Sahar, La Fazia Vincenzo Fazia, Ströker Erwin, Gianni Carola, Zeriouh Sarah, Bianchi Stefano, Sieira Juan, Combes Stephane, Sarkozy Andrea, Rossi Pietro, Boveda Serge, Natale Andrea, de Asmundis Carlo, Chierchia Gian-Battista, HRMC Investigators. Pulsed electric field, cryoballoon, and radiofrequency for paroxysmal atrial fibrillation ablation: a propensity score-matched comparison.. Europace : European pacing, arrhythmias, and cardiac electrophysiology : journal of the working groups on cardiac pacing, arrhythmias, and cardiac cellular electrophysiology of the European Society of Cardiology. 2023. DOI: 10.1093/europace/euae016.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление