Comparative evaluation of Vitek®2 and broth microdilution method for colistin susceptibility testing of Gram-negative isolates from intensive care unit in a tertiary care hospital.

Ananda Thripthi, Vandana K E, Mukhopadhyay Chiranjay

СтатьяСравнительное исследованиеОценочное исследованиеПоддержка службы здравоохранения СШАПеревод пока не подготовлен

Аннотация

INTRODUCTION

Colistin is the last resort treatment against resistant Gram-negative bacteria, necessitating reliable and rapid means for sensitivity testing of colistin. Automated systems like VITEK®2 are adopted to determine the minimum inhibitory concentration (MIC) due to easy usage. Broth microdilution (BMD) for colistin MIC was suggested by EUCAST and CLSI.

OBJECTIVE

To compare and evaluate colistin MIC by BMD and VITEK®2 against Gram-negative organisms from the ICU in a tertiary care hospital.

METHOD: Clinically significant organisms isolated from ICU patients were included. MIC was determined using BMD and VITEK®2. Very major error (VME), major error (ME), essential agreement (EA), categorical agreement (CA), positive predictive value (PPV), negative predictive value (NPV), sensitivity, and specificity were analysed.

RESULT: 533 isolates were obtained from blood (435,81.60%), respiratory samples (57,10.70%), pus and exudates (20,3.80%), urine (18,3.40%), and CSF (3,0.60%). The Enterobacterales were K. pneumoniae (185,34.70%) E. coli (73,13.70%) and E. cloacae (26,4.90%) while non-fermenters were A. baumannii (209,39.20%) and P. aeruginosa (40,7.50%). The VITEK®2 sensitivity was >99%; specificity ranged from 14.28 to 52.94%. PPV was 93.81% while NPV was 93.75%. VME ranged from 47 to 100% between isolates. ME was up to 20%. The highest VME was obtained in E. coli (100%). The total EA and CA observed were 68.5% and 99.79% respectively.

CONCLUSION

Automated system VITEK®2 failed to detect the resistance in 32 (60%) isolates. The obtained VME and ME values were >3%, which is unacceptable as per the standard guidelines. EA of ≥90% wasn't obtained. Sensitivity for VITEK®2 was >99%, but had low specificity (14.28%). Hence, VITEK®2 is not reliable for colistin susceptibility testing.

Для цитирования: Ananda Thripthi, Vandana K E, Mukhopadhyay Chiranjay. Comparative evaluation of Vitek®2 and broth microdilution method for colistin susceptibility testing of Gram-negative isolates from intensive care unit in a tertiary care hospital.. Indian journal of medical microbiology. 2024. DOI: 10.1016/j.ijmmb.2024.100559.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.