Comparison of techniques for identification of peripheral vestibular nystagmus.
Аннотация
OBJECTIVE
To determine the best clinical method for identifying peripheral vestibular nystagmus, by comparing eye movement examination with optic fixation, and with fixation removed using Frenzel's glasses, infra-red video-Frenzel's goggles or an ophthalmoscope, with results of electronystagmography.
METHOD: One hundred patients referred for electronystagmography from the audiovestibular medicine clinic at Queen Alexandra Hospital, Portsmouth, were examined immediately before undergoing electronystagmography.
RESULTS
Video-Frenzel's goggles were highly effective at detecting peripheral vestibular nystagmus, with a sensitivity of 85 per cent (95 per cent confidence interval, 62.1-96.8 per cent) and a specificity of 65 per cent (53.5-75.3 per cent), compared with electronystagmography. Ophthalmoscopy had comparable sensitivity to Frenzel's glasses (used in the dark), i.e. 26.3 per cent (9.1-51.2 per cent) compared with 31.6 per cent (12.6-56.6 per cent), respectively. Frenzel's glasses as normally used in ENT clinics (i.e. in dim lighting) were ineffective, with a sensitivity of just 10 per cent (1.2-31.7 per cent).
CONCLUSION
Video-Frenzel's goggles should be used in all clinics with substantial numbers of balance-impaired patients. Traditional Frenzel's glasses have no place in clinical practice unless formal black-out facilities are available.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.