DOI 10.1016/j.jcrs.2013.11.024

Медицинская библиотека · Journal of cataract and refractive surgery · 2014

Intraocular polyimide intraocular lens haptic breakage long-term postoperatively.

Stallings Shannon, Werner Liliana, Chayet Arturo, Masket Samuel, Camacho Fidel, Cutler Peck Carolee, Mamalis Nick

Клинический случайСтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

UNLABELLED: We present 2 cases of 3-piece silicone intraoculanbsp;r lenses (IOLs) with broken polyimide haptics. In the first case, the IOL was implanted in the anterior chamber after posterior capsule rupture. Twelve years later, it was explanted and exchanged because of bullous keratopathy. However, on careful manipulation for explantation, the haptics shattered into multiple pieces. In the second case, initial in-the-bag implantation of the IOL was uneventful. Fifteen years later, the IOL dislocated into the anterior chamber as a result of spontaneous in situ haptic breakage. The haptics of both IOLs appeared very brittle during careful manipulation for gross and light microscopy, with further breakage. Scanning electron microscopy of both IOLs revealed no evidence of haptic degradation. Their surfaces appeared smooth and regular with jagged edges where the breaks occurred. Few studies have examined the long-term biocompatibility of polyimide, and more research is needed to determine the cause of this phenomenon.

FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.

Для цитирования: Stallings Shannon, Werner Liliana, Chayet Arturo, Masket Samuel, Camacho Fidel, Cutler Peck Carolee, Mamalis Nick. Intraocular polyimide intraocular lens haptic breakage long-term postoperatively.. Journal of cataract and refractive surgery. 2014. DOI: 10.1016/j.jcrs.2013.11.024.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.