Clinical evaluation of the Altostar HDV RT-PCR Kit 1.5.

Eichholz Julia Carolin, Bremer Birgit, Wedemeyer Heiner, Maasoumy Benjamin, Sandmann Lisa

СтатьяОценочное исследованиеПеревод в очереди

Аннотация

BACKGROUND

Detection and quantification of hepatitis D virus RNA is essential for diagnosis and treatment of chronic hepatitis D virus (HDV) infection. This study evaluated the diagnostic performance of the Altostar HDV RT-PCR Kit 1.5 in real-world clinical samples with and without treatment with bulevirtide (BLV).

METHODS

Plasma samples of 40 patients with chronic HDV infection were analyzed with the Altostar HDV RT-PCR Kit 1.5 and compared to the local standard, the RoboGene HDV RNA Quantification Kit 2.0, with additional analyses of corresponding serum and plasma samples and on-treatment samples during BLV therapy.

RESULTS

HDV RNA levels in plasma samples quantified by both assays showed a significant positive correlation. All samples were quantifiable by Altostar 1.5, whereas two samples were undetectable by Robogene 2.0 During BLV therapy, mean baseline (BL) and follow-up (FU) HDV RNA levels were comparable between the two assays. There was a similar proportion of patients with virological response (Altostar 1.5: 75% [n = 15/20] vs. Robogene 2.0: 70% [n = 14/20]), whereas the proportion of samples with undetectable HDV RNA at FU was considerably higher in measurements with Robogene 1.5 (40% vs. 0%).

CONCLUSIONS

HDV RNA levels quantified by the Altostar HDV RT-PCR Kit 1.5 were comparable to the local standard. The higher number of detectable samples may reflect increased sensitivity, occasional false-positive results with Altostar 1.5, or the impact of automated extraction in the Robogene workflow. Importantly, treatment response during BLV treatment was not influenced by the type of quantification assay.

Для цитирования: Eichholz Julia Carolin, Bremer Birgit, Wedemeyer Heiner, Maasoumy Benjamin, Sandmann Lisa. Clinical evaluation of the Altostar HDV RT-PCR Kit 1.5.. Journal of clinical virology : the official publication of the Pan American Society for Clinical Virology. 2026. DOI: 10.1016/j.jcv.2026.105971.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.