Acute type A aortic dissection complicated by aortic stent graft collapse.

Szabolcs Zoltan, Hüttl Kalman, Laczko Agnes, Daroczi Laszlo, Hüttl Tivadar, Paulovich Erzsebet, Hartyanszky Istvan

Клинический случайСтатьяПеревод пока не подготовлен

Аннотация

A 57-year-old man complaining of chest pain presented with signs of lower limb ischemia 1 year after implantation of a stent graft at the aortoiliac bifurcation. A computed tomography scan revealed the presence of a type A aortic dissection and complete collapse of the stent graft by bulging of the false lumen. The patient underwent emergency surgical reconstruction of the aortic root and arch, which allowed reexpansion of the previously collapsed stent graft. Stenting of residual stenoses distal to the stent graft and of an occluded left renal artery was also successful.