DOI 10.1016/j.thromres.2014.09.024

Медицинская библиотека · Thrombosis research · 2014

Towards improved diagnosis of von Willebrand disease: comparative evaluations of several automated von Willebrand factor antigen and activity assays.

Favaloro Emmanuel J, Mohammed Soma

Сравнительное исследованиеОценочное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

INTRODUCTION

von Willebrand disease (VWD) is reportedly the most common bleeding disorder and arises from deficiency and/or defects of von Willebrand factor (VWF). Laboratory diagnosis and typing has important management implications and requires a wide range of tests, including VWF activity and antigen, and involves differential identification of qualitative vs quantitative defects.

METHODS

We have assessed several VWF antigen and activity assays (collagen binding [VWF:CB], ristocetin cofactor [VWF:RCo] and the new Siemens INNOVANCE assay [VWF:Ac], employing latex particles and gain of function recombinant glycoprotein Ib to facilitate VWF binding and agglutination without need for ristocetin) using different instrumentation, including the new Sysmex CS-5100, with a large sample test set (n=600). We included retrospective plus prospective study designs, and also evaluated desmopressin responsiveness plus differential sensitivity to high molecular weight VWF.

RESULTS

VWF:Ag and VWF:RCo results from different methods were respectively largely comparable, although some notable differences were evident, including one high false normal VWF:Ag value (105 U/dL) on a type 3 VWD sample, possibly due to heterophile antibody interference in the latex-based CS-5100 methodology. VWF:Ac was largely comparable to VWF:RCo, but VWF:CB showed discrepant findings to both VWF:RCo and VWF:Ac with some patients, most notably patients with type 2M VWD.

CONCLUSIONS

(a) VWF:Ag on different platforms are largely interchangeable, as are VWF:RCo on different platforms, except for occasional (some potentially important) differences, and manufacturer recommended methods may otherwise require some assay optimization; (b) VWF:RCo and VWF:Ac are largely interchangeable, except for occasional differences that may also relate to assay design (differing optimizations); (c) VWF:CB provides an additional activity to supplement VWF:RCo or VWF:Ac activity assays, and is not interchangeable with either.

Для цитирования: Favaloro Emmanuel J, Mohammed Soma. Towards improved diagnosis of von Willebrand disease: comparative evaluations of several automated von Willebrand factor antigen and activity assays.. Thrombosis research. 2014. DOI: 10.1016/j.thromres.2014.09.024.