[Bruch's Membrane Opening in OCT in Correlation with Corneal Correction - a Pilot Study].
Аннотация
UNLABELLED:
Background
The measurement by optical coherence tomography (OCT) of the Bruch membrane opening (BMO) and the thinnest retinal fiber nerve layer in relation to it (BMO-MRW) has been performed in clinical routine since 2014. To compensate for astigmatism, or increased or decreased corneal power, a correction is performed through the mean K-value of the measured eye. The aim of this study was to measure the value of this correction and its influence on the measurement results.
Methods
The sectors of BMO-MRW and BMO values of the five right eyes of five healthy patients were measured five times each with Heidelberg Spectralis OCT. Corneal compensation was systematically raised with each single measurement (7.1, 7.4, 7.7, 8.0, 8.3 mm).
RESULTS
The data showed almost linear dependence on the given corneal compensation values, with intraindividual variability. For the BMO-MRW, only small effects of compensation were found (0.85 up to 1.97 % per K-value difference of 0.3 mm). For BMO, the effect was greater, with a mean change of 7.71 % for every 0.3 mm change in compensation.
Conclusion
For BMO-MRW, corneal compensation is of low clinical relevance. BMO is more dependent on this correction. In follow-up measurements, the compensation might not account for significant changes, although we recommend using correct corneal compensation values when obtaining single or first-time measurements.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.