MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.jcrs.2012.12.035

Медицинская библиотека · Journal of cataract and refractive surgery · 2013

Histomorphological and immunofluorescence evaluation of clear corneal incisions after microcoaxial phacoemulsification with 2.2 mm and 1.8 mm systems.

Vasavada Abhay R, Johar Kaid, Praveen Mamidipudi R, Vasavada Viraj A, Arora Anshul I

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

PURPOSE: To compare changes in the incision's histomorphology and denaturation of collagen I in rabbit eyes having microcoaxial phacoemulsification through 2.2 mm and 1.8 mm incision-compatible systems.

DESIGN

Randomized experimental trial.

SETTING

Iladevi Cataract & IOL Research Centre, Ahmedabad, India.

METHODS

Thirty rabbit eyes were randomized into Group 1 (microcoaxial phacoemulsification through 2.2 mm incisions using Infiniti system [torsional ultrasound]) and Group 2 (microcoaxial phacoemulsification through 1.8 mm incisions using Stellaris system [longitudinal ultrasound]). Each group was then divided into 3 subgroups of 5 eyes each based on 1 of the 3 intervention options: phacoemulsification only, intraocular lens (IOL) insertion only, and phacoemulsification with IOL insertion. Left eyes were randomized for microcoaxial phacoemulsification, and right eyes were treated as controls.

RESULTS

After phacoemulsification, eyes in Group 1 showed loss of epithelium at the roof of the incisions and Descemet membrane detachment at the floor of the incisions. These findings did not change after IOL insertion. After phacoemulsification, eyes in Group 2 showed loss of epithelium, but Descemet membrane remained attached. There was a longitudinal split in the incision's stroma in the direction of internal entry. The stromal damage increased after IOL implantation. Immunofluorescence studies showed no obvious irregularities in the arrangement of collagen I in either group. A dot blot analysis showed significant denaturation of collagen I in Group 2.

CONCLUSION

The histomorphology of the 2.2 mm system incision showed localized Descemet membrane detachment and endothelial cell loss. The 1.8 mm system incision showed exaggerated stromal damage after IOL insertion.

Для цитирования: Vasavada Abhay R, Johar Kaid, Praveen Mamidipudi R, Vasavada Viraj A, Arora Anshul I. Histomorphological and immunofluorescence evaluation of clear corneal incisions after microcoaxial phacoemulsification with 2.2 mm and 1.8 mm systems.. Journal of cataract and refractive surgery. 2013. DOI: 10.1016/j.jcrs.2012.12.035.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.