DOI 10.1016/j.cmpb.2023.107862
A simplified coronary model for diagnosis of ischemia-causing coronary stenosis.
Аннотация
BACKGROUND AND
OBJECTIVE
The functional assessment of the severity of coronary stenosis from coronary computed tomography angiography (CCTA)-derived fractional flow reserve (FFR) has recently attracted interest. However, existing algorithms run at high computational cost. Therefore, this study proposes a fast calculation method of FFR for the diagnosis of ischemia-causing coronary stenosis.
METHODS
We combined CCTA and machine learning to develop a simplified single-vessel coronary model for rapid calculation of FFR. First, a zero-dimensional model of single-vessel coronary was established based on CCTA, and microcirculation resistance was determined through the relationship between coronary pressure and flow. In addition, a coronary stenosis model based on machine learning was introduced to determine stenosis resistance. Computational FFR (cFFR) was then obtained by combining the zero-dimensional model and the stenosis model with inlet boundary conditions for resting (cFFRr) and hyperemic (cFFRh) aortic pressure, respectively. We retrospectively analyzed 75 patients who underwent clinically invasive FFR (iFFR), and verified the model accuracy by comparison of cFFR with iFFR.
RESULTS
The average computing time of cFFR was less than 2 s. The correlations between cFFRr and cFFRh with iFFR were r = 0.89 (p < 0.001) and r = 0.90 (p < 0.001), respectively. Diagnostic accuracy, sensitivity, specificity, positive predictive value, negative predictive value, positive likelihood ratio, negative likelihood ratio for cFFRr and cFFRh were 90.7%, 95.0%, 89.1%, 76.0%, 98.0%, 8.7, 0.1 and 92.0%, 95.0%, 90.9%, 79.2%, 98.0%, 10.5, 0.1, respectively.
CONCLUSIONS
The proposed model enables rapid prediction of cFFR and exhibits high diagnostic performance in selected patient cohorts. The model thus provides an accurate and time-efficient computational tool to detect ischemia-causing stenosis and assist with clinical decision-making.
Предметные рубрики MeSH
Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Ошибки и ограничения
История · было → стало
Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.
Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.