MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1007/bf00464627

Медицинская библиотека · Archives of oto-rhino-laryngology · 1988

Click ABR intensity-latency characteristics in diagnosing conductive and cochlear hearing losses.

Steinhoff H J, Böhnke F, Janssen T

СтатьяОбзорПеревод пока не подготовлен

Аннотация

A schematic description of the correlation between various pathologies of hearing impairments and the behavior of auditory brainstem responses (ABR) is presented. Conductive pathology and high-frequency cochlear hearing losses prolong wave component latency due to energy loss and hair cell dysfunction. In cases of flat cochlear hearing loss latency is not affected. Prolonged interwave latencies between wave I and wave V indicate eight nerve and brainstem disorders. An algorithm was developed in the form of a flow chart for locating various malfunctions. Fields of wave V intensity-latency functions were designed for the faster detection and more precise evaluation of conductive and cochlear hearing losses.

Для цитирования: Steinhoff H J, Böhnke F, Janssen T. Click ABR intensity-latency characteristics in diagnosing conductive and cochlear hearing losses.. Archives of oto-rhino-laryngology. 1988. DOI: 10.1007/bf00464627.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.