MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1097/mao.0000000000003628

Медицинская библиотека · Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology · 2022

Promontory Electrocochleography Recordings to Predict Speech-Perception Performance in Cochlear Implant Recipients.

Walia Amit, Shew Matthew A, Lee David S, Lefler Shannon M, Kallogjeri Dorina, Wick Cameron C, Durakovic Nedim, Fitzpatrick Douglas C, Ortmann Amanda J, Herzog Jacques A, Buchman Craig A

СтатьяПоддержка NIH: внешняя программаПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

OBJECTIVE

1) To determine the relationship of electrocochleography (ECochG) responses measured on the promontory with responses measured at the round window (RW) and various intracochlear sites. 2) To evaluate if promontory ECochG responses correlate with postoperative speech-perception performance using the cochlear implant (CI).

STUDY

DESIGN

Prospective cohort study.

SETTING

Tertiary referral center.

PATIENTS AND INTERVENTIONS: Ninety-six adult CI recipients with no cochlear malformations or previous otologic surgery.

MAIN OUTCOME MEASURES: Acoustically evoked ECochG responses were measured intraoperatively at both extracochlear and intracochlear locations. ECochG total response (ECochG-TR), a measure of residual cochlear function, was calculated by summing the fast Fourier transformation amplitudes in response to 250-Hz to 2-kHz acoustic stimuli. Speech-perception performance was measured at 3 months.

RESULTS

There were strong linear correlations for promontory ECochG-TR with the ECochG-TRs measured at the RW ( r = 0.95), just inside scala tympani ( r = 0.91), and after full insertion ( r = 0.83). For an individual subject, the morphology of the ECochG response was similar in character across all positions; however, the response amplitude increased from promontory to RW (~1.6-fold) to just inside scala tympani (~2.6-fold), with the largest response at full insertion (~13.1-fold). Promontory ECochG-TR independently explained 51.8% of the variability ( r2 ) in consonant-nucleus-consonant at 3 months.

CONCLUSIONS

Promontory ECochG recordings are strongly correlated with responses previously recorded at extracochlear and intracochlear sites and explain a substantial portion of the variability in CI performance. These findings are a critical step in supporting translation of transtympanic ECochG into the clinic preoperatively to help predict postoperative CI performance.

Для цитирования: Walia Amit, Shew Matthew A, Lee David S, Lefler Shannon M, Kallogjeri Dorina, Wick Cameron C, Durakovic Nedim, Fitzpatrick Douglas C, Ortmann Amanda J, Herzog Jacques A, Buchman Craig A. Promontory Electrocochleography Recordings to Predict Speech-Perception Performance in Cochlear Implant Recipients.. Otology & neurotology : official publication of the American Otological Society, American Neurotology Society [and] European Academy of Otology and Neurotology. 2022. DOI: 10.1097/mao.0000000000003628.