DOI 10.1515/cclm-2024-0373

Медицинская библиотека · Clinical chemistry and laboratory medicine · 2024

Insight into the status of plasma renin and aldosterone measurement: findings from 526 clinical laboratories in China.

Zhou Weiyan, Deng Yuhang, Ma Wenjun, Zhao Haijian, Wang Kaijun, Zhang Qian, Gan Wei, Chen Wenxiang, Cai Jun, Zhang Chuanbao

СтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

Accurate measurements of renin and aldosterone levels play an important role in primary aldosteronism screening, which is of great importance in the management and categorization of hypertension. The objective of this study is to investigate the current status of plasma renin and aldosterone measurements in China, which is achieved by analyzing the results of 526 clinical laboratories nationwide for three pooled fresh plasma samples derived from more than 2,000 patients.

METHODS

Renin and aldosterone in three pooled plasma samples were measured four times in 526 laboratories employing various measurement systems. The inter- and intra-laboratory %CV were calculated and compared. To determine the source of the substantial inter-laboratory %CV, laboratories were categorized according to the measurement systems they are using, and both the inter- and intra-measurement-system %CV were calculated and compared.

RESULTS

Regarding renin, the majority of laboratories use four primary commercial immunoassays. However, for aldosterone, in addition to commercial immunoassays, laboratory-developed liquid chromatography-tandem mass spectrometry (LC-MS) methods are also used by laboratories. The median values of intra-laboratory %CVs, intra-measurement-system %CVs, inter-laboratory %CVs, and inter-measurement systems %CVs varied between 1.6 and 2.6 %, 4.6 and 14.9 %, 8.3 and 25.7 %, and 10.0 and 34.4 % for renin, respectively. For aldosterone, these values ranged from 1.4 to 2.2 %, 2.5-14.7 %, 9.9-31.0 %, and 10.0-35.5 %, respectively.

CONCLUSIONS

The precision within laboratories and measurement systems for plasma renin and aldosterone measurements is satisfactory. However, the comparability between laboratories using different measurement systems remains lacking, indicating the long way to achieve standardization and harmonization for these two analytes.

Для цитирования: Zhou Weiyan, Deng Yuhang, Ma Wenjun, Zhao Haijian, Wang Kaijun, Zhang Qian, Gan Wei, Chen Wenxiang, Cai Jun, Zhang Chuanbao. Insight into the status of plasma renin and aldosterone measurement: findings from 526 clinical laboratories in China.. Clinical chemistry and laboratory medicine. 2024. DOI: 10.1515/cclm-2024-0373.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.