DOI 10.1111/hae.70115

Медицинская библиотека · Haemophilia : the official journal of the World Federation of Hemophilia · 2025

Drug-Specific Calibration: A Solution to Reagent Variability in Extended Half-Life Factor IX Activity Measurements.

Grand François, Blanc-Jouvan Florence, Mourey Guillaume, Ryman Anne, Eschwege Valérie, Brionne-François Marie, Lasne Dominique, Voisin Sophie, Le Cam-Duchez Véronique, Diaz-Cau Isabelle, Flaujac Claire, Noyel Pauline, Lavenu-Bombled Cecile, Demaistre Emmanuel, Jeanpierre Emmanuelle, Donnard Magali, Pouplard Claire, Hezard Nathalie, Sattler Laurent, Nougier Christophe

СтатьяМногоцентровое исследованиеПеревод пока не подготовлен

Аннотация

INTRODUCTION

Discrepancies in factor IX activity (FIX:C) measurements between one-stage clotting assays (OSAs) have been observed following infusion with recombinant factor IX extended half-life concentrates (EHL-rFIX) in the treatment of haemophilia B. These variations, primarily due to differences in activated partial thromboplastin time (APTT) reagents, complicate clinical decision-making.

OBJECTIVES

The aim of this study was to evaluate whether drug-specific calibrations for albumin-fused recombinant FIX (rFIX-FP) and Fc-fused recombinant FIX (rFIX-Fc) could reduce inter-reagent discrepancies.

METHODS

In a multicentre field study involving 20 laboratories, plasma samples spiked with rFIX-FP, rFIX-Fc or standard rFIX were tested using different APTT reagents. FIX:C was measured with usual and drug-specific calibration. Data were analysed for compliance (target range: 80%-120% of expected values), interlaboratory variability and intralaboratory reproducibility.

RESULTS

Usual calibration resulted in significant discrepancies among reagents, with compliance rates varying widely. Drug-specific calibration significantly improved compliance for all reagents tested, except for one concentration with rFIX-Fc. Interlaboratory variability decreased markedly, with coefficients of variation dropping from 19.4%-36.0% (usual calibration) to 6.0%-12.4% (specific calibration). Intralaboratory reproducibility was consistent whatever the type of calibration.

CONCLUSION

Drug-specific calibration for EHL-rFIX reduces reagent-related variability in OSA FIX assays, ensuring reliable and standardised results. This approach facilitates monitoring with a single reagent system, improving laboratory efficiency. Wider availability of validated calibrators remains crucial for broader adoption and standardisation.

Для цитирования: Grand François, Blanc-Jouvan Florence, Mourey Guillaume, Ryman Anne, Eschwege Valérie, Brionne-François Marie, Lasne Dominique, Voisin Sophie, Le Cam-Duchez Véronique, Diaz-Cau Isabelle, Flaujac Claire, Noyel Pauline, Lavenu-Bombled Cecile, Demaistre Emmanuel, Jeanpierre Emmanuelle, Donnard Magali, Pouplard Claire, Hezard Nathalie, Sattler Laurent, Nougier Christophe. Drug-Specific Calibration: A Solution to Reagent Variability in Extended Half-Life Factor IX Activity Measurements.. Haemophilia : the official journal of the World Federation of Hemophilia. 2025. DOI: 10.1111/hae.70115.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.