DOI 10.1097/00003246-199702000-00005

Медицинская библиотека · Critical care medicine · 1997

Intratracheal pressure monitoring during synchronized intermittent mandatory ventilation and pressure controlled-inverse ratio ventilation.

Warters R D, Allen S J, Tonnesen A S

Сравнительное исследованиеСтатьяПеревод пока не подготовлен

Аннотация

OBJECTIVES

To directly measure airway pressures proximal and distal to endotracheal tubes during conventional synchronized intermittent mandatory ventilation (SIMV) and pressure controlled-inverse ratio ventilation (PC-IRV), and to compare them with these values measured by the ventilator.

DESIGN

Prospective, nonrandomized study.

SETTING

Surgical intensive care unit at a trauma center.

PATIENTS: Group 1: Eight intubated adult patients connected to mechanical ventilators in the SIMV mode were studied. All patients required mechanical ventilation following traumatic injuries. Group 2: Five intubated adult patients with adult respiratory distress syndrome connected to mechanical ventilators were studied.

INTERVENTIONS: A small polyethylene catheter was threaded through each endotracheal tube such that it could be positioned to measure pressures proximal and distal to the tubes.

MEASUREMENTS AND MAIN

RESULTS

During SIMV, a significant pressure gradient exists across endotracheal tubes. In addition, although initiation of PC-IRV did lead to a lower peak airway pressure measured proximally, intratracheal peak airway pressure was unchanged.

CONCLUSIONS

A pressure gradient exists during inspiration from the ventilator to the trachea in mechanically ventilated patients. Tracheal pressures cannot be predicted from proximal airway pressure monitors because of marked variation in endotracheal tube resistance in vivo. Initiation of PC-IRV does not result in a decrease in peak airway pressure when measured intratracheally.

Для цитирования: Warters R D, Allen S J, Tonnesen A S. Intratracheal pressure monitoring during synchronized intermittent mandatory ventilation and pressure controlled-inverse ratio ventilation.. Critical care medicine. 1997. DOI: 10.1097/00003246-199702000-00005.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.