MedElevator Войти в исследования
← Поиск публикаций

DOI 10.1016/j.hrthm.2020.12.031

Медицинская библиотека · Heart rhythm · 2021

Intermediate-term performance and safety of His-bundle pacing leads: A single-center experience.

Teigeler Todd, Kolominsky Jeffrey, Vo Chau, Shepard Richard K, Kalahasty Gautham, Kron Jordana, Huizar Jose F, Kaszala Karoly, Tan Alex Y, Koneru Jayanthi N, Ellenbogen Kenneth A, Padala Santosh K

СтатьяПоддержка исследования: вне правительства СШАПеревод пока не подготовлен

Аннотация

BACKGROUND

The short-term safety, feasibility, and performance of His-bundle pacing (HBP) leads have been reported; however, their longer-term performance beyond 1 year remains unclear.

OBJECTIVE

The purpose of this study was to examine the intermediate-term performance and safety of HBP.

METHODS

All HBP lead implants at Virginia Commonwealth University between January 2014 and January 2019 were analyzed. HBP was performed using a Medtronic SelectSecure 3830-69 cm pacing lead.

RESULTS

Of 295 attempts, successful HBP implantation (selective or nonselective) was seen in 274 cases (93%). Mean follow-up duration was 22.8 ± 19.5 months (median 19.5; interquartile range 11-33). Mean age was 69 ± 15 years; 58% were males; and ejection fraction <50% was noted in 30%. Indications for pacemaker included sick sinus syndrome in 41%, atrioventricular block in 36%, cardiac resynchronization therapy in 7%, and refractory atrial fibrillation in 15%. Selective HBP was achieved in 33%. Mean HBP capture threshold at implant was 1.1 ± 0.9 V at 0.8 ± 0.2 ms, which significantly increased at chronic follow-up to 1.7 ± 1.1 V at 0.8 ± 0.3 ms (P <.001). Threshold was ≥2.5 V in 24% of patients, and 28% had an increase in HBP threshold ≥1 V. Loss of His-bundle capture at follow-up (septal right ventricular pacing) was seen in 17%. There was a total of 31 (11%) lead revisions, primarily for unacceptably high thresholds.

CONCLUSION

Although HBP can prevent or improve pacing-induced cardiomyopathy, the elevated capture thresholds, loss of His-bundle capture, and lead revision rates at intermediate follow-up are of concern. Longer-term follow-up data from multiple centers are needed.

Для цитирования: Teigeler Todd, Kolominsky Jeffrey, Vo Chau, Shepard Richard K, Kalahasty Gautham, Kron Jordana, Huizar Jose F, Kaszala Karoly, Tan Alex Y, Koneru Jayanthi N, Ellenbogen Kenneth A, Padala Santosh K. Intermediate-term performance and safety of His-bundle pacing leads: A single-center experience.. Heart rhythm. 2021. DOI: 10.1016/j.hrthm.2020.12.031.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.