Efficacy of the current of injury in envisaging the dislodgement of leads implanted in the right atrial septum or the right ventricular septum.

Yoshiyama Tomotaka, Shimeno Kenji, Matsuo Masanori, Matsumoto Ryo, Matsumura Yoshiki, Abe Yukio, Naruko Takahiko, Yoshiyama Minoru

СтатьяНаблюдательное исследованиеПеревод пока не подготовлен

Аннотация

BACKGROUND

The implantation of leads in the right atrial septum (RAS) or the right ventricular septum (RVS) is technically challenging, and dislodgement occurs occasionally. This study aims to determine a predictor for the dislodgement of leads implanted in the RAS or RVS.

METHODS

This retrospective cohort study enrolled 137 consecutive patients who underwent the cardiac implantable electronic devices implantation, using active fixation leads in the RAS and RVS. We compared the pacing threshold, R- or P-wave amplitude, slew rate, and presence of the current of injury (COI) between dislodged and nondislodged leads.

RESULTS

We performed lead fixation for 74 and 125 times in the RAS and RVS, respectively. Atrial lead dislodgement occurred five times (6.8%) intraoperatively and five times (6.8%) postoperatively, whereas ventricular lead dislodgement occurred eight times (6.4%) intraoperatively and three times (2.4%) postoperatively. Although there were no lead parameters that showed a significant difference common to RAS lead and RVS lead, the presence of the COI was significantly different between nondislodged and dislodged leads in both the RAS and RVS (atrial leads: 57.8% vs 0%, P < 0.001; ventricular leads: 67.5% vs 9.1%, P < 0.001). The positive predictive value of COI presence for predicting no lead dislodgement was 100% and 98.7% in the RAS and RVS, respectively.

CONCLUSION

Lead dislodgement is more likely when the COI is absent; documentation of COI should be pursued during lead implantation in challenging sites as the RAS and RVS.

Для цитирования: Yoshiyama Tomotaka, Shimeno Kenji, Matsuo Masanori, Matsumoto Ryo, Matsumura Yoshiki, Abe Yukio, Naruko Takahiko, Yoshiyama Minoru. Efficacy of the current of injury in envisaging the dislodgement of leads implanted in the right atrial septum or the right ventricular septum.. Pacing and clinical electrophysiology : PACE. 2019. DOI: 10.1111/pace.13666.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.