PURPOSE: Despite the use of steroid-eluting leads, a transient but not persistent rise in the atrial/ventricular capture threshold (TRACT/TRVCT) can occur early after pacemaker implantation in patients with sick sinus syndrome. This study aimed to assess the prevalence, predictors, and mechanisms of TRACT/TRVCT in patients with heart failure undergoing implantable cardioverter defibrillator (ICD) or cardiac resynchronization therapy (CRT) implantation.

METHOD: One hundred twenty consecutive patients underwent ICD (N = 70) or CRT (N = 50) implantation. Capture threshold was measured at implantation, 7-day, 1-month, and 6-month post-implantation. TRACT/TRVCT was defined as a threshold rise at 7 days by more than twice the height of the threshold at implantation, with full recovery during follow-up. Atrial and brain natriuretic peptide (ANP and BNP) levels were measured before implantation.

RESULTS

TRACT and TRVCT were observed in 13 (11%) and 10 (8%) patients, respectively. Patients with TRACT had lower ANP level (median 72 [42-105] vs. 99 [49-198] pg/mL, P = 0.06), lower ANP/BNP ratio (0.29 [0.20-0.36] vs. 0.50 [0.33-0.70], P < 0.01), lower atrial sensing amplitude (2.0 ± 0.8 vs. 2.7 ± 1.3 mV, P = 0.02), and lower left ventricular ejection fraction (32 ± 12 vs. 40 ± 14%, P = 0.04) than those without TRACT. TRACT recovered within 1 month, whereas TRVCT recovered within 6 months. In multivariable analysis, ANP/BNP ratio was the only independent predictor of TRACT (OR, 0.018; 95% CI, 0.001-0.734; P = 0.034).

CONCLUSIONS

Atrial degenerative change characterized by lower ANP/BNP ratio was associated with the occurrence of TRACT in patients with heart failure. TRVCT could also occur, but it required a longer recovery time than TRACT.

Для цитирования: Ogawa Kojiro, Yoshida Kentaro, Uehara Yoshiko, Ebine Mari, Kimata Akira, Nishina Hidetaka, Takeyasu Noriyuki, Noguchi Yuichi, Ieda Masaki, Aonuma Kazutaka, Nogami Akihiko. Mechanistic implication of decreased plasma atrial natriuretic peptide level for transient rise in the atrial capture threshold early after ICD or CRT-D implantation.. Journal of interventional cardiac electrophysiology : an international journal of arrhythmias and pacing. 2018. DOI: 10.1007/s10840-018-0409-0.

Предметные рубрики MeSH

Подготовка и перевод публикации
Обработка статьи · этапы, ошибки и управление
Загрузка состояния…

История · было → стало

Состояние обновляется каждые 5 секунд, пока блок открыт. Процент показывается только там, где известен объём работы. Пауза останавливает продолжение этапа; текущий запрос к источнику может завершиться.

Тип публикации указан по данным источника. Уровень доказательности автоматически не присваивается.